12.6 General Specifications

Item

Specifications

 

 

Reference junction compensation accuracy

 

 

 

Above 0°C with input terminal temperature balanced (60 minutes after power on)

 

Type R, S, B, W, WRe: ±1.0°C

 

 

Type K, J, E, T, N, L, U: ±0.5°C

 

Maximum input voltage

±10 VDC (continuous) for ranges of 200 mV or less, TC, RTD, and DI ranges

 

±60 VDC (continuous) for 2 VDC or higher ranges

Input resistance

Approximately 10 Mor more for ranges of 200 mV or less and TC

 

Approximately 1 Mfor 2 VDC or higher ranges

Input source resistance

Volt, TC:

2 kor less

 

 

RTD input:

10 or less per wire (The resistance of all three wires must be equal).

Bias current

10 nA or less (except when burnout detection function is enabled)

Common mode rejection ratio 120 dB (50/60 Hz ± 0.1%, 500 unbalanced, between the minus terminal and ground)

Normal mode rejection ratio

40 dB or more (50/60 Hz ± 0.1%)

Noise rejection

By the integrating A/D converter: The integration time is 16.7 ms, 20 ms, or 100 ms (dot model)

 

Low-pass filtering (pen model): Filter time constant is 2 s, 5 s, or 10 s

 

Moving average (dot model):

The number of samples is 2 to 16.

 

 

 

 

Effects of Operating Conditions

Item

Specifications

 

 

Ambient temperature

With temperature variation of 10°C:

 

Measurement: ±(0.1% of rdg + 1 digit) or less

 

Recording:

Measurement fluctuation + ±0.2% of the recording span or less

 

* Excluding the error of reference junction compensation

Power supply fluctuation

With variation within 90 to 132 V and 180 to 264 VAC (50/60 Hz):

 

Measurement: ±1 digit or less

 

Recording:

±0.1% of the recording span or less

 

With variation of ±2 Hz from rated power frequency (power supply voltage 90 to 132, 180 to 264 VAC):

 

Measurement: ±(0.1% of rdg + 1 digit) or less

 

Recording:

Same as the measurement fluctuation

Magnetic field

AC (50/60 Hz) and DC 400 A/m fields:

 

Measurement: ±(0.1% of rdg 10 digit) or less

 

Recording:

±0.5% of the recording span or less

Input source resistance

 

 

 

DC voltage range

With variation of +1 k:

 

 

Ranges of 200 mV or less: Within ±10 V

 

Ranges of 2V or more:

±0.1% of rdg

TC range

With variation of +1 k: ±10 V or less

RTD range

With variation of 10 per wire (resistance of all three wires must be equal):

 

Measurement: ±(0.1% of rdg + 1 digit) or less

 

Recording:

Measurement fluctuation + ±0.1% of the recording span or less

 

With maximum difference of 40 mbetween wires: Approx. 0.1°C (for Pt100)

Mounting position

With a backward tilt of 30° or less

 

Measurement: ±(0.1% of rdg + 1 digit) or less

 

Recording:

Indication fluctuation + ±0.1% of the recording span or less

Vibration

When a sinusoidal vibration along all three axis at a frequency between 10 to 60 Hz and an

 

acceleration of 0.2 m/s2 is applied for 2 hours

 

Measurement: ±(0.1% of rdg + 1 digit) or less

 

Recording:

Indication fluctuation + ±0.1% of the recording span or less

 

 

 

 

Other Specifications

 

Item

Specifications

 

Clock

With a calendar function

 

 

 

 

Accuracy of clock

±100 ppm, excluding a delay (of 1 second, maximum) caused each time the power is turned on.

 

 

 

 

DST

The date/time for switching between standard time and DST can be specified.

 

 

 

 

Memory backup

A built-in lithium battery backs up the setup parameters and runs the clock

 

 

(battery life: approximately ten years at room temperature).

 

 

 

 

Key lock function

Password for releasing the key lock can be set.

 

Key lock items

RCD key, MENU key, DISP MENU key, FEED key, and function under the FUNC key (Alarm ACK, computation start/

 

 

stop, print, message, printout buffer clear, periodic printout reset, and pen or ribbon cassette exchange)

 

 

 

 

Internal light

Lights up the recording area of the chart paper. Uses a dedicated LED.

 

 

 

 

Noise

Machine Noise Information Ordinance 3.GSGV, Jan 18, 1991

 

 

Maximum noise level: 60 dB (A) or less (complies with ISO7779)

 

 

 

 

 

 

12-20

Page 222
Image 222
Omega RD100B manual Effects of Operating Conditions, Other Specifications, 12-20