Chapter 6 Setup Operations for Convenient Functions (Setting Mode)

6.1Setting the Trend Recording Interval (Dot Model).

This section explains the details of setting the trend recording interval (dot printing interval) on the dot model.

Procedure

1.Hold down the MENU key for 3 seconds to enter Setting mode.

2.Press the key or SHIFT + key to select Aux and then press the key.

3.Press the key or SHIFT + key to select Trend and then press the key.

4.Set the dot printing interval and press the key.

Use the key or SHIFT + key to select values.

For the procedure on how to enter values or characters, see section 4.2.

If you press the ESC key in the middle of the operation, the settings entered up to then are cancelled, and the display returns to a higher level menu.

5.When the Setting complete screen appears, do either of the following: To correct the setting, press the key.

If you are done, press the ESC key.

6.Hold down the MENU key for 3 seconds to return to Operation mode.

6

Set=Aux

Aux=Trend

Mode=Auto

Select the trend recording interval mode.

Setup Operations for Convenient

ESC/?

Trend Interval

The new setting takes effect.

Setting complete

 

Explanation

Mode

Auto: The trend recording interval is set according to the chart speed in the range of 10 s to 90 s so that the dots do not overlap. The interval is displayed on the System display (see page 12-9).

Fix: The trend recording interval is fixed to 10 s/6 dots. On models with the computation function (/M1 option), the dot printing interval varies depending on the number of measurement and computation channels to be trend recorded.

Number of channels

Trend recording

to be recorded

interval

 

 

6 or less

10 s

7 to 12

15 s

13 to 18

20 s

Dot

When set to Auto

When set to Fix

Functions (Setting Mode)

6-1

Page 109
Image 109