5.1 Setting the Input range

Delta Computation

Set=Range

CH=02-02

First channel Last channel

Mode=Delta

Ref.CH=01

Span_L= -2.000

Span_R= 2.000

02-02 Channel ESC/? Setting complete

Description

Set the channel range.

Select Delta.

Set the reference channel.

Set the left span value.

Set the right span value.

The new setting takes effect.

Reference Channel

The reference channel must be a channel that is smaller in channel number than the channel being set.

Span Left and Span Right

The range type is set to the same type as the reference channel. The span left and span right can be set in the range shown below. Span left and span right cannot be set to the same value.

Input Type

Range Type

Selectable Span Range (°C)

Selectable Span Range (°F)

DC voltage

20 mV

–20.00 to 20.00 mV

 

(Volt)

60 mV

–60.00 to 60.00 mV

 

 

200 mV

–200.0 to 200.0 mV

 

 

2 V

–2.000 to 2.000 V

 

 

6 V

–6.000 to 6.000 V

 

 

20 V

–20.00 to 20.00 V

 

 

50 V

–50.00 to 50.00 V

 

 

 

 

 

Thermocouple

R

1760.0 to 1760.0 °C

–3168 to 3168 °F

(TC)

S

–1760.0 to 1760.0 °C

–3168 to 3168 °F

 

B

–1820.0 to 1820.0 °C

–3276 to 3276 °F

 

K

–1570.0 to 1570.0 °C

–2826 to 2826 °F

 

E

–1000.0 to 1000.0 °C

–1800 to 1800 °F

 

J

–1300.0 to 1300.0 °C

–1999.9 to 2340.0 °F

 

T

–600.0 to 600.0 °C

–1080.0 to 1080.0 °F

 

N

–1300.0 to 1300.0 °C

–2340 to 2340 °F

 

W

–1999.9 to 2315.0 °C

–4167 to 4167 °F

 

L

–1100.0 to 1100.0 °C

–1980 to 1980 °F

 

U

–600.0 to 600.0 °C

–1080.0 to 1080.0 °F

 

WRe

–1999.9 to 2400.0 °C

–4320 to 4320 °F

 

 

 

 

RTD

PT (Pt100)

–800.0 to 800.0 °C

–1440.0 to 1440.0 °F

 

JPT (JPt100)

–750.0 to 750.0 °C

–1350.0 to 1350.0 °F

 

 

 

 

5-4