Epson XDT manual Viewing Serial Number Mode, SetUp State

Page 32

Section 3: Instrument

The instrument will display the measurement using the newly acquired calibration values.

The instrument computes the dewpoint from the measured capacitance of the sensor. The sequence of computations is as follows:

The capacitance is converted to percent of full scale (-100°C to +20°C dp), using the low dB, low dew, and saturated dB when performing SpanCheck values.

Cal Adj is applied to correct the mid point of the curve.

If enabled, the multi-point calibration table is applied to further correct the transfer function.

If enabled, the single point calibration is applied to fine tune the transfer function.

The corrected percent of full scale is converted to dewpoint.

3.4.4.4 Viewing Serial Number Mode

In this mode the user can view the instrument serial number. The display shows the serial number by alternately displaying XSN and the number. If the number is larger than 1999 then it is dis-

played in 2 parts, first part is the thousands signified by the x103 legend in the upper right corner of the display and the second part is the units. For example serial number 12345 will be shown as:

x10 3

Pressing the UP and DOWN buttons simultaneously resets the instrument, this is useful in instal- lations where power can not be turned on and off easily. Pressing the Pressure Correct button will cause the instrument to enter the setup state.

Pressing the MODE button, returns the unit to the Viewing Dewpoint Mode.

3.4.5 SetUp State

To enter the Setup State power-up the unit (or reset it from the serial number Mode), and hold the MODE key pressed while the unit is initializing. Refer to Appendix B for a flow diagram of the SetUp State.

The setup state provides eleven capabilities, each one controlled from its own mode. Depressing the MODE button navigates through the modes in the following order: nSelect Alternate Display Units oSelect Sensor Type pSet Measured Attenuation of Sensor at a Low Dewpoint qSet the Low Dewpoint at which the Attenuation was measured rSet the Calibration Mid-range Adjust sAnalog Output select 4/20mA or 0/24mA tPerform a Test and Verification of the Analog Out- put uSet dewpoint corresponding to low end of Analog Output vSet dewpoint corresponding to high end of Analog Output wLock/Unlock the instrument nnSensor Cable Compensation (back to) n.

These eleven functions are explained in their respective sections which follow. Note that modes 6,7,8&9 will appear only on instruments with the analog output option. Changes can be made only to an unlocked instrument, the changes are stored (when the MODE button is pressed) and retained even if the instrument power is turned off.

1.Display of alternate units: In this mode, a second engineering unit can be chosen to be dis- played alternating with the engineering unit selected in the operating state, for example, a dewpoint can be alternately shown in °C and in PPM.

25

Image 32
Contents Yyyyyyy y Sensor Serial No Instrument Serial No XDT User’s Manual Iii XDT User’s Manual Table of Contents Index Overview of the XDT XDT User’s Manual Sensor and Sampling Techniques Precautions using the sensor Sensor Technical Specifications Sensor Installation & Sampling TechniquesIn-situ Installation Not RecommendedExtractive Installation Mechanical Installation Troubleshooting unexpected readings SATInstrument Precautions using the XDT Electromagnetic Compatibility ConsiderationsInstrument Technical Specifications IsolationInstallation Instrument Mechanical InstallationXDT-PM DIN43700 Enclosure Installation XDT-NEMA IP65 Enclosure InstallationConnecting Power Electrical ConnectionsAC Mains Electrical Power Connection Low Voltage DC Powered Option Electrical Power ConnectionWiring the Alarm Contacts Sensor ConnectionInterfacing to the Analog Output Starting up Operating the InstrumentInterfacing to the RS-232 option Push Buttons Display Conventions12 3 4 5 6 7 8 Viewing Dewpoint Mode Operating StateXDT User’s Manual Alarms HI **** ˚C Start CalibrationSpanCheck Mode Single Point Self Calibration, manual or scheduled Instrument XDT User’s Manual Viewing Serial Number Mode SetUp StateXDT User’s Manual Secured Password Protection Resetable Audio-Visual Alarm Option Nfpa compliant Troubleshooting the Instrument Display Explanation Required Action Maintenance XDT User’s Manual Glossary DB9 Glossary XDT User’s Manual Appendix a Flow Diagram of Operating State User Interface Sign on message Calibration without key pressHere Appendix B Flow Diagram of Set-Up State User Interface 16.24˚C˚FAppendix Sensor MechanicalAppendix D Optional Sensor Fittings Appendix E XDT Circuit Board Dimensions Side ViewAppendix E XDT-PM Enclosure Mounting Dimensions Appendix E XDT-NEMA Enclosure Mounting Dimensions Appendix F Sensor/SpanCheck Theory of Operation Appendix G Dewpoint Response time Analysis XDT User’s Manual Appendices Appendix H Sample Gas Filter Considerations Tested Instrument ˚CAppendix I XDT Grounding Considerations SensorAppendix J Analog Output vs. Dewpoint Instrument reading in ppmV Appendix K RS-232C Interface Protocol Help commandReceive Help Data Who Reset Login Data command Error SensOpen or Error SensShort or Error SensSat K n m? command Serial Number CommandWho Command Login CommandAppendix L Procedure for Exchanging XDT Sensors State User InterfaceAppendix M Uncertainty in LBS & ppmV calculations Xentaur Return Authorization Number Request Form To obtain a RAN fill out this form and fax to 631Page Index Enclosure dimensions Pounds of H2O per million standard cubic Voltage, mains supply Iii Warranty