Epson XDT manual Appendix a Flow Diagram of Operating State User Interface

Page 44

XDT User’s Manual Appendices

Appendix A: Flow Diagram of Operating State User Interface

Power ON initialization

30 seconds anywhere (other than

appears when changing

and sign on message

"Calibration") without key press

settings of locked unit

 

 

 

HI/LO

˚C

 

 

 

 

and/orHIFlashingLO

correspondingindicates

energizednotrelays

DEWPOINT PSI

 

{

 

 

 

 

units

 

 

 

 

SET

 

 

 

 

HI

 

 

1)

 

{

 

 

(Relay

 

 

 

 

 

choices

 

 

HIAlarm

 

SET

 

 

 

{

 

 

 

 

HI

˚C

 

 

 

DEWPOINT

 

installed

Setting

 

values

 

 

 

 

 

 

 

 

SET

 

 

 

 

HI

˚C

options are

 

 

 

 

 

{

 

alarm

 

 

values

 

 

 

SET

 

if

 

 

LO

 

(Relay2)

 

 

 

Displayedonly

 

{

 

 

LOAlarm

 

choices

 

 

 

{

 

 

 

 

SET

 

 

 

 

LO

˚C

 

 

 

DEWPOINT

 

 

Setting

 

values

 

 

 

 

 

 

 

 

SET

 

 

 

 

LO

˚C

 

 

 

 

 

 

 

{

 

values

Toggle Pressure Correction

Viewing Dewpoint

Flashing PSI indicates pressure corrected dewpoint

SHORT PRESS

LONG PRESS

AND PSI MODE

Set activity options

Set trigger dewpoint

Set hysteresis

Set activity options

Set trigger dewpoint

Set hysteresis

1/2 sec

 

Toggle 14.7 psi

SET

 

VALUE

View/Set

PSI

Sensor

 

{

Pressure

SHORT PRESS

values

 

 

LONG PRESS

SET

Toggle 14.7 psi

 

VALUE

View/Set

PSI

Gas

 

{

Pressure

SHORT PRESS

values

LONG PRESS

 

Relay polarity:

= energize relay above trigger dewpoint

= energize relay below trigger dewpoint appropriate polarity selection in conjunction with proper wiring wll provide fail safe operation

Scroll through Alarm activity options

Fault reaction:

= faults do Not effect relays

= faults "look" like High dewpoint

= faults "look" like Low dewpoint = faults Flash the relay

Cancel Calibration (it is not performed)

When a third Alarm Relay is installed,

 

 

 

 

 

 

 

setup menu options will appear here

 

 

 

 

 

 

 

similar to Relays 1&2, but with both

 

 

 

 

 

 

 

the HI&LO legends active. This is

 

 

 

 

 

 

 

also the dewpoint alarm for units with

 

 

 

 

 

 

 

resetable Audio-Visual Alarm.

 

 

 

 

 

 

 

SET

few sec

SET

NUMBERS

 

Sensor

SET

 

Start

 

CALCULATING

 

 

 

Calibration

 

CALIBRATION

OK

Reading

 

 

 

 

 

 

see Self-Cal user

 

 

 

TOO LOW

Confirm

{

 

 

 

 

 

 

 

 

 

 

 

SpanCheck

interface next page

 

 

 

 

 

 

 

 

 

 

again

 

 

 

 

 

 

 

 

Confirm SpanCheck

 

 

 

 

 

 

 

Viewing

 

SET

1 sec

 

SET

SET

 

Serial-

go to

 

 

 

 

 

 

 

 

 

 

 

number

Set-Up

 

 

 

 

1 minute

{

 

State

 

 

 

 

 

 

 

 

 

 

 

Press together

 

New calibration

 

ignore error

 

to RESET the

 

values are saved

 

 

 

 

 

 

 

 

 

instrument

 

 

 

 

 

 

Legend: A slash '' in the display area, is used to depict two alternately shown (flashing back and forth) messages. denotes beeping. Buttons without designator or arrow leading out, perform no function

37

Image 44
Contents Yyyyyyy y Sensor Serial No Instrument Serial No XDT User’s Manual Iii XDT User’s Manual Table of Contents Index Overview of the XDT XDT User’s Manual Sensor and Sampling Techniques Precautions using the sensor Sensor Technical Specifications Sensor Installation & Sampling TechniquesIn-situ Installation Not RecommendedExtractive Installation Mechanical Installation Troubleshooting unexpected readings SATInstrument Precautions using the XDT Electromagnetic Compatibility ConsiderationsInstrument Technical Specifications IsolationInstallation Instrument Mechanical InstallationXDT-PM DIN43700 Enclosure Installation XDT-NEMA IP65 Enclosure InstallationConnecting Power Electrical ConnectionsAC Mains Electrical Power Connection Low Voltage DC Powered Option Electrical Power ConnectionWiring the Alarm Contacts Sensor ConnectionInterfacing to the Analog Output Starting up Operating the InstrumentInterfacing to the RS-232 option Push Buttons Display Conventions12 3 4 5 6 7 8 Viewing Dewpoint Mode Operating StateXDT User’s Manual Alarms HI **** ˚C Start CalibrationSpanCheck Mode Single Point Self Calibration, manual or scheduled Instrument XDT User’s Manual Viewing Serial Number Mode SetUp StateXDT User’s Manual Secured Password Protection Resetable Audio-Visual Alarm Option Nfpa compliant Troubleshooting the Instrument Display Explanation Required Action Maintenance XDT User’s Manual Glossary DB9 Glossary XDT User’s Manual Appendix a Flow Diagram of Operating State User Interface Sign on message Calibration without key pressHere Appendix B Flow Diagram of Set-Up State User Interface 16.24˚C˚FAppendix Sensor MechanicalAppendix D Optional Sensor Fittings Appendix E XDT Circuit Board Dimensions Side ViewAppendix E XDT-PM Enclosure Mounting Dimensions Appendix E XDT-NEMA Enclosure Mounting Dimensions Appendix F Sensor/SpanCheck Theory of Operation Appendix G Dewpoint Response time Analysis XDT User’s Manual Appendices Appendix H Sample Gas Filter Considerations Tested Instrument ˚CAppendix I XDT Grounding Considerations SensorAppendix J Analog Output vs. Dewpoint Instrument reading in ppmV Appendix K RS-232C Interface Protocol Help commandReceive Help Data Who Reset Login Data command Error SensOpen or Error SensShort or Error SensSat K n m? command Serial Number CommandWho Command Login CommandAppendix L Procedure for Exchanging XDT Sensors State User InterfaceAppendix M Uncertainty in LBS & ppmV calculations Xentaur Return Authorization Number Request Form To obtain a RAN fill out this form and fax to 631Page Index Enclosure dimensions Pounds of H2O per million standard cubic Voltage, mains supply Iii Warranty