Epson manual XDT User’s Manual

Page 33

XDT User’s Manual

CAUTION:

2), 3), 4) and 5) are set at the factory and need only be modified when a sensor is changed.

2. Selecting the sensor type: In this mode the user can select the software matching the type of

sensor installed in the instrument: XT1 : XTR-100 (-100°C to +20°C); XT6 : XTR-65 (-65°C to +20°C);

DB : show impedance of any sensor, disable Open and Short error massages.

3&4Adjusting low end sensor attenuation and dewpoint: These modes are used to enter a data pair representing a low dewpoint and the sensor attenuation measured at this low dewpoint. (see Sensor Theory of Operation appendix F). This data pair should not be modified unless the sensor is replaced (refer to procedures for replacing a sensor in appendix L). The sensor ship- ping container is labeled with the proper values.

5.Calibration Adjustment: In this mode the user can enter a sensor specific Adjustment Value to improve the instruments’ accuracy in the range of -50°C to -10°C, this adjustment has very little or no effect outside this range. The Calibration Adjustment Value is printed on the ship- ping container supplied with the sensor, or may be derived if the sensor can be exposed to a known reference.

Deriving an Adjustment Value: If the adjustment value for your sensor is not known, contact your representative and follow the instructions below for entering known values. If the value is not available, then it may be derived using a manual method, provided that a moisture cali- bration facility is available.

The calibration must be done in the range of -25°C to -20°C (attempting to calibrate outside this range may cause inaccuracies), the dewpoint must be kept stable during calibration and it must be measured accurately by a reference instrument such as a chilled mirror. One must also note the state of pressure compensation and turn it on or off as applicable to the reference. The necessary steps are described below (also refer to the Set-Up State flow diagram in Appendix B):

a.Make sure that the low attenuation and low dewpoint are correctly entered for this sensor.

b.SpanCheck™ the sensor.

c.Dry-down the sensor for at least 12 hrs.

d.Expose the sensor to the known dewpoint (-25°C to -20°C), for a sufficient time (at least 30 min.)

e.Go to the CAL/ADJ mode and press the UP button, the display will show the current value, in the range of -1.99 to +1.99 followed by blinking horizontal lines.

f.Observe the display, if the buttons are not being pressed; the display will every few seconds show for a short duration the dewpoint as being computed at the moment (utilizing the current cal-adjust value in the computation).

g.Use the UP or DOWN buttons to modify the cal-adjust value and observe the alternately displayed dewpoint. Perform this adjustment until the dewpoint matches the value shown by the reference instrument.

h.Press the MODE button to go to the next mode, this will save the new cal-adjust value, and it will be retained even if power is turned off.

Entering a known Adjustment Value: If the sensor is being replaced an adjustment value should be supplied with the sensor, to enter it into the instrument:

26

Image 33
Contents Yyyyyyy y Sensor Serial No Instrument Serial No XDT User’s Manual Iii XDT User’s Manual Table of Contents Index Overview of the XDT XDT User’s Manual Sensor and Sampling Techniques Precautions using the sensor Sensor Installation & Sampling Techniques Sensor Technical SpecificationsNot Recommended In-situ InstallationExtractive Installation Mechanical Installation SAT Troubleshooting unexpected readingsElectromagnetic Compatibility Considerations Instrument Precautions using the XDTIsolation Instrument Technical SpecificationsInstrument Mechanical Installation InstallationXDT-PM DIN43700 Enclosure Installation XDT-NEMA IP65 Enclosure InstallationElectrical Connections Connecting PowerLow Voltage DC Powered Option Electrical Power Connection AC Mains Electrical Power ConnectionWiring the Alarm Contacts Sensor ConnectionInterfacing to the Analog Output Operating the Instrument Interfacing to the RS-232 optionStarting up Display Conventions 12 3 4 5 6 7 8Push Buttons Operating State Viewing Dewpoint ModeXDT User’s Manual Alarms Start Calibration HI **** ˚CSpanCheck Mode Single Point Self Calibration, manual or scheduled Instrument XDT User’s Manual SetUp State Viewing Serial Number ModeXDT User’s Manual Secured Password Protection Resetable Audio-Visual Alarm Option Nfpa compliant Troubleshooting the Instrument Display Explanation Required Action Maintenance XDT User’s Manual Glossary DB9 Glossary XDT User’s Manual Sign on message Calibration without key press Appendix a Flow Diagram of Operating State User InterfaceHere 16.24˚C˚F Appendix B Flow Diagram of Set-Up State User InterfaceSensor Mechanical AppendixAppendix D Optional Sensor Fittings Side View Appendix E XDT Circuit Board DimensionsAppendix E XDT-PM Enclosure Mounting Dimensions Appendix E XDT-NEMA Enclosure Mounting Dimensions Appendix F Sensor/SpanCheck Theory of Operation Appendix G Dewpoint Response time Analysis XDT User’s Manual Appendices Tested Instrument ˚C Appendix H Sample Gas Filter ConsiderationsSensor Appendix I XDT Grounding ConsiderationsAppendix J Analog Output vs. Dewpoint Instrument reading in ppmV Help command Receive Help Data Who Reset Login Data commandAppendix K RS-232C Interface Protocol K n m Error SensOpen or Error SensShort or Error SensSatSerial Number Command ? commandWho Command Login CommandState User Interface Appendix L Procedure for Exchanging XDT SensorsAppendix M Uncertainty in LBS & ppmV calculations To obtain a RAN fill out this form and fax to 631 Xentaur Return Authorization Number Request FormPage Index Enclosure dimensions Pounds of H2O per million standard cubic Voltage, mains supply Iii Warranty