Epson XDT manual SpanCheck Mode

Page 28

Section 3: Instrument

3.4.4.3.1 SpanCheck™ Mode

The instrument is calibrated at the factory with the sensor it is shipped with and does not need to be re-calibrated prior to installation.

Instrument calibration is recommended in approximately 12 month intervals, and the XTR65W sensor should be re-calibrated after prolonged exposure to liquid water. Simply follow steps 1 - 7 of the procedure below, removing the sensor from the sample gas stream. To avoid injury, make sure the gas stream is depressurized before removing the sensor.

It is recommended to keep the sensor exposure to room air as short as possible, in order to avoid super saturation of the sensor. While super saturation is not damaging to the sensor, it will prolong the initial dry-down time after you install the sensor in the sample stream. Therefore, remove the sensor from the packaging container only after you are ready to proceed with the calibration pro- cedure and install the sensor in the sample stream immediately after the calibration procedure is completed. If you are not ready to use the sensor right away after calibration, put the sensor back in the shipping container for dry storage.

The instrument must be calibrated with the sensor it will be used with.The calibration procedure takes advantage of the sensor’s ability to saturate (refer to Appendix F) and is executed by the instrument computer, by performing the following steps:

If a single point calibration is in effect with a reference dewpoint below -70°C, it should be dis- abled (see next section) before performing SpanCheck calibration.

1.If the instrument is locked and a calibration is attempted; it will display LOC and will not per- form the calibration. To unlock the instrument consult section 3.4.5.-10

2.Push the MODE key a few times until the display shows CAL.

3.Press the UP button. The display will show CNF, prompting you to confirm that you want to start the calibration procedure. You can abort the calibration procedure by pressing the MODE key.

4.Remove the sensor from its packaging container or sample stream, so that the porous metal filter is visible and the sensor is exposed to in-hand micro-climate (refer to explanation below). Close the packaging container as soon as you have removed the sensor to avoid degra- dation of the desiccant inside the container. You may want to re-use the container at a later date.

5.Expose the sensor to in-handmicro-climate, and push the UP button again to confirm that you want to start the calibration procedure. The display will flash AC for 60 seconds, while the sen- sor is saturating. Make sure you keep the sensor exposed to in-handmicro-climate until the display shows END.

6.After 60 seconds, the display will flash the selected sensor type (see Selecting a sensor type 3.4.5.-2)and then the instrument will calculate the slope and offset of the sensor curve while displaying the calculations and then will display END for a few seconds, after which the instru- ment will automatically go into measuring mode. The sensor calibration is completed. (The display may flash SAT, indicating that the sensor has super-saturated. As soon as the sensor is exposed to an atmosphere with a dewpoint lower than the saturation dewpoint, the display will indicate the dewpoint measured by the sensor.) The display may show alternating SEN and 2LO as an indication that the measured capacitance is too low to be from a saturated sensor, in this case make sure that the sensor is properly plugged in, and repeat the calibration procedure.

7.Install the sensor in the sample cell or adapter fitting or put it back into the packaging con- tainer for later use.

21

Image 28
Contents Yyyyyyy y Sensor Serial No Instrument Serial No XDT User’s Manual Iii XDT User’s Manual Table of Contents Index Overview of the XDT XDT User’s Manual Sensor and Sampling Techniques Precautions using the sensor Sensor Technical Specifications Sensor Installation & Sampling TechniquesIn-situ Installation Not RecommendedExtractive Installation Mechanical Installation Troubleshooting unexpected readings SATInstrument Precautions using the XDT Electromagnetic Compatibility ConsiderationsInstrument Technical Specifications IsolationInstallation Instrument Mechanical InstallationXDT-PM DIN43700 Enclosure Installation XDT-NEMA IP65 Enclosure InstallationConnecting Power Electrical ConnectionsAC Mains Electrical Power Connection Low Voltage DC Powered Option Electrical Power ConnectionWiring the Alarm Contacts Sensor ConnectionInterfacing to the Analog Output Interfacing to the RS-232 option Operating the InstrumentStarting up 12 3 4 5 6 7 8 Display ConventionsPush Buttons Viewing Dewpoint Mode Operating StateXDT User’s Manual Alarms HI **** ˚C Start CalibrationSpanCheck Mode Single Point Self Calibration, manual or scheduled Instrument XDT User’s Manual Viewing Serial Number Mode SetUp StateXDT User’s Manual Secured Password Protection Resetable Audio-Visual Alarm Option Nfpa compliant Troubleshooting the Instrument Display Explanation Required Action Maintenance XDT User’s Manual Glossary DB9 Glossary XDT User’s Manual Appendix a Flow Diagram of Operating State User Interface Sign on message Calibration without key pressHere Appendix B Flow Diagram of Set-Up State User Interface 16.24˚C˚FAppendix Sensor MechanicalAppendix D Optional Sensor Fittings Appendix E XDT Circuit Board Dimensions Side ViewAppendix E XDT-PM Enclosure Mounting Dimensions Appendix E XDT-NEMA Enclosure Mounting Dimensions Appendix F Sensor/SpanCheck Theory of Operation Appendix G Dewpoint Response time Analysis XDT User’s Manual Appendices Appendix H Sample Gas Filter Considerations Tested Instrument ˚CAppendix I XDT Grounding Considerations SensorAppendix J Analog Output vs. Dewpoint Instrument reading in ppmV Receive Help Data Who Reset Login Data command Help commandAppendix K RS-232C Interface Protocol Error SensOpen or Error SensShort or Error SensSat K n m? command Serial Number CommandWho Command Login CommandAppendix L Procedure for Exchanging XDT Sensors State User InterfaceAppendix M Uncertainty in LBS & ppmV calculations Xentaur Return Authorization Number Request Form To obtain a RAN fill out this form and fax to 631Page Index Enclosure dimensions Pounds of H2O per million standard cubic Voltage, mains supply Iii Warranty