XDT User’s Manual
ber can be corrected by reversing the use of the UP and DOWN keys), with the “12” flashing. Pressing the “MODE” button will flash the next “0”. Pressing the “UP” button three times will display “1230”. Pressing the “MODE” button will flash the last “0”. Pressing the “UP” button four times will display “1234”. Pressing and holding the “MODE” button, will display “SET UN/ LOC”, indicating that the instrument is now unlocked. It may be locked without the password, same as any other instrument.
Note: To avoid unauthorized changes, it is imperative that instrument is returned into the locked mode.
11.Sensor Cable Compensation: As noted in other paragraphs this instrument measures the impedance of the sensor (in the form of attenuation in decibels) from which it computes the dewpoint. At very low dewpoints the sensor capacitance is small enough to be in the same order of magnitude as a long coaxial cable. Thus the measurement circuit must be compen- sated for the impedance of the sensor cable. Instruments which are delivered with a cable are already compensated at the factory and should not be compensated again, however if
anew cable is connected, the following procedure should be used for compensation (also refer to the
a.Disconnect the sensor from the cable by unplugging the BNC connector, keep the cable connected to the instrument.
b.Prepare a means by which the BNC connector at the end of the cable could be reliably shorted (using a jumper clip etc.), do not attempt to short at the instrument screw terminals as this approach will not account for the cable resistance.
c.Go to the CNF/CBL/OPN (Confirm Cable Open) mode.
d.Leave the BNC connector open and press the UP key. The instrument will show OPN and beep for a few seconds. Then it will show CNF/CBL/SHR (Confirm Cable Short).
e.Short the BNC connector and press the UP key, the unit will show SHR and beep for a few seconds. Then it will show CNF/CBL/OPN again.
f.Leave the BNC connector open again and press the UP button. The instrument will show OPN and beep for a few seconds. Then it will show END
The compensation data will be retained even if power is turned off.
3.5 Resetable Audio-Visual Alarm Option (NFPA compliant)
When the instrument is ordered with an
•The alarm set point is programmed on a third alarm, hence the HI and LO alarm relays are free to be used for additional purposes The third alarm is properly configured to 39°F at the fac- tory, but when necessary may be modified by the user. Refer to the notes following HI & LO Alarms in the Flow Diagram of Operating State User Interface in appendix A.
•The visual indicator is the display backlight. When there is no alarm condition the LCD dis- play is not backlit but can still be easily and clearly read (in an extremely dark room the back- light may be momentarily turned on by pressing the TEST button). However as long as an alarm condition exists the LCD will be backlit in red. This
28