Epson XDT manual Display Explanation Required Action

Page 37

XDT User’s Manual

 

DISPLAY

EXPLANATION

 

 

REQUIRED ACTION

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LO

ERR



PROM check sum failed.

tests

 

 

 

 

 

 

 

 

HI

ERR



RAM write/read test failed.

 

cycle power

on

 

 

 

 

 

 

 

%

ERR



Unidentified power-up failure.

power

 

if problem persists,

 

 

 

 

 

 

C.S.F/



EEPROM Check Sum Failed.

 

return to your representative

 

 

 

 

 

 

 

 

ERR/AD



A/D converter failure.

2 min.

 

for service.

 

 

 

 

 

ERR/REF



Reference voltage for A/D out of spec.

 

 

 

 

 

 

per

 

 

LO/BAT



Low supply voltage.

 

 

once

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LO RNG/



Instrument low temperature range

tests

 

 

TEMP

 

has been exceeded.

 

make sure that the unit is at a

 

system

 

 

 

 

 

 

 

HI

RNG/



Instrument high temperature range

 

temperature of -10°C to +50°C.

TEMP

 

has been exceeded.

 

 

 

 

 

 

 

 

 

OPN/



Sensor circuit is open.

dewpoint display

 

 

 

 

 

 

 

SAT/



Sensor is saturated.

tests

see table in sensor troubleshooting section 2.5

 

 

 

 

SHR/



Sensor circuit is shorted.

 

 

 

 

 

 

...

 

 

Trying to calculate dewpoint for undefined sensor.

 

select sensor and recalibrate.

 

 

 

 

 

 

 

 

 

 

 

DB

 



Trying to calibrate an undefined sensor.

calibration

 

 

 

 

 

 

tests

see calibration instructions.

SEN/2LO



Sensor reading is ‘too’ low to be from a saturated sensor,

 

 

 

 

thus can not be calibrated

 

 

 

 

 

 

 

 

2LO

 

 

While Compensating for cable length, the measurement did not

cable

tests

Check the means of opening & shorting the sensor

 

 

 

correspond to the expected open or short values.

cable. see section 3.4.5 - 11

2HI

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ERR/EEP



EEPROM write cycle not completed.

 

 

if this persists, return for service.

 

 

 

 

 

 

 

LOC

 



Attempting to modify a locked unit.

 

 

unlock unit, see set-up state 3.4.5.-10.

 

 

 

 

 

 

 

SEC

 



Attempting to unlock a secured unit.

 

 

 

 

 

 

 

 

 

 

HI

(flashing

while

The HI Alarm (alarm #1) relay is deenergized.

 

 

See section 3.4.4.2 Setting the Alarms.

viewing DEWPOINT)

 

 

 

 

 

 

 

 

 

 

 

LO

(flashing

while

The LO Alarm (alarm #2) relay is deenergized.

 

 

 

viewing DEWPOINT)

 

miscellaneous

 

 

 

 

 

 

 

A.O./OPN



The Analog Output ‘sees’ an open circuit

 

check the cables connecting the analog output, make

 

 

 

 

 

sure the terminating impedance is less that 500Ω.

 

 

 

 

 

 

PSI flashing when dis-

Pressure correction is enabled, the dewpoint is computed using

 

refer to section 3.4.4.1

 

 

playing dewpoint

 

programmed pressures

 

 

 

 

 

 

 

 

-70°C (-94°F) with

An XTR-65 sensor is measuring less than -70°C or -94°F D.P.

 

 

check instrument

flashing DEWPOINT

 

 

 

upgrade to XTR-100 sensor

 

 

 

 

 

 

RNG

 

number can not be displayed in 3½ digits

 

 

 

 

 

 

 

 

 

 

XEN

 



Turn on message, Xentaur (Greek Ξ=X)

 

 

 

 

 

 

 

 

display is dark

No power to instrument

 

 

check power to instrument, check fuse

 

 

 

 

 

 

 

Legend:  denotes a beeping accompaniment to the message.

/denotes alternately flashing messages.

30

Image 37
Contents Yyyyyyy y Sensor Serial No Instrument Serial No XDT User’s Manual Iii XDT User’s Manual Table of Contents Index Overview of the XDT XDT User’s Manual Sensor and Sampling Techniques Precautions using the sensor Sensor Installation & Sampling Techniques Sensor Technical SpecificationsNot Recommended In-situ InstallationExtractive Installation Mechanical Installation SAT Troubleshooting unexpected readingsElectromagnetic Compatibility Considerations Instrument Precautions using the XDTIsolation Instrument Technical SpecificationsInstrument Mechanical Installation InstallationXDT-PM DIN43700 Enclosure Installation XDT-NEMA IP65 Enclosure InstallationElectrical Connections Connecting PowerLow Voltage DC Powered Option Electrical Power Connection AC Mains Electrical Power ConnectionWiring the Alarm Contacts Sensor ConnectionInterfacing to the Analog Output Interfacing to the RS-232 option Operating the InstrumentStarting up 12 3 4 5 6 7 8 Display ConventionsPush Buttons Operating State Viewing Dewpoint ModeXDT User’s Manual Alarms Start Calibration HI **** ˚CSpanCheck Mode Single Point Self Calibration, manual or scheduled Instrument XDT User’s Manual SetUp State Viewing Serial Number ModeXDT User’s Manual Secured Password Protection Resetable Audio-Visual Alarm Option Nfpa compliant Troubleshooting the Instrument Display Explanation Required Action Maintenance XDT User’s Manual Glossary DB9 Glossary XDT User’s Manual Sign on message Calibration without key press Appendix a Flow Diagram of Operating State User InterfaceHere 16.24˚C˚F Appendix B Flow Diagram of Set-Up State User InterfaceSensor Mechanical AppendixAppendix D Optional Sensor Fittings Side View Appendix E XDT Circuit Board DimensionsAppendix E XDT-PM Enclosure Mounting Dimensions Appendix E XDT-NEMA Enclosure Mounting Dimensions Appendix F Sensor/SpanCheck Theory of Operation Appendix G Dewpoint Response time Analysis XDT User’s Manual Appendices Tested Instrument ˚C Appendix H Sample Gas Filter ConsiderationsSensor Appendix I XDT Grounding ConsiderationsAppendix J Analog Output vs. Dewpoint Instrument reading in ppmV Receive Help Data Who Reset Login Data command Help commandAppendix K RS-232C Interface Protocol K n m Error SensOpen or Error SensShort or Error SensSatSerial Number Command ? commandWho Command Login CommandState User Interface Appendix L Procedure for Exchanging XDT SensorsAppendix M Uncertainty in LBS & ppmV calculations To obtain a RAN fill out this form and fax to 631 Xentaur Return Authorization Number Request FormPage Index Enclosure dimensions Pounds of H2O per million standard cubic Voltage, mains supply Iii Warranty