Epson manual Appendix L Procedure for Exchanging XDT Sensors, State User Interface

Page 62

XDT User’s Manual Appendices

Appendix L: Procedure for Exchanging XDT Sensors

To exchange sensors one must know the measured low attenuation of the sensor at some low dew- point as well as the calibration adjustment value. These numbers can be found on the shipping container serial number label.

Through the following steps it is helpful to refer to Appendix A: Flow Diagram of XDT Set-Up

State User Interface

1.Power up (or reset) the XDT while holding down the MODE key, in order to enter the Set-Up state. Make sure that the instrument in unlocked, refer to section 3.4.5.-10.

2.Press the MODE key to go to the “Choose Sensor Type” mode. Make sure that the proper sensor type is chosen, e.g. XT1 (for XTR-100) or XT6 (for XTR-65).

3.Press the MODE key to go to the “Set Measured Attenuation @ low dewpoint” mode. The display will show the currently entered attenuation in decibels (dB), this is indicated by hav- ing both the °C and °F on the display. Use the UP or DOWN keys to select the value which appears on the shipping container label. Note that this value is a negative number typically in the range of -18.0DB to -14.0DB.

4.Press the MODE key to go to the “Set low dewpoint” mode. The display will show the cur- rently entered low dewpoint. Use the UP or DOWN keys to select the value which appears on the shipping container label.

5.Press the MODE key to go to the “Calibration Midrange Adjust” mode. The display will alternately show CAL and ADJ. Press the UP key once to see the entered calibration adjustment value. The display will alternately show the currently entered value and the dewpoint being measured, when showing the Cal Adj value the number (in the range of -2.0 to +2.0) will be followed by a blinking “X”, and when showing the dewpoint the display will have the appro- priate legend e.g. °C. Use the UP or DOWN keys to select the value which appears on the shipping container label.

6.Press the MODE key five (5) times to go to the “Lock or Unlock” mode. The display will alternately show UN and LOC. Press the DOWN key to lock the instrument. The display will alternately show only LOC.This prevents inadvertent changes.

7.Press the MODE key once.

8.Now the instrument has stored in it’s EEPROM, the calibration values associated with the new sensor, HOWEVER NOTE THAT IT IS STILL NOT CALIBRATED. The instrument may now be reset.

9.Perform the SpanCheck Procedure in section 3.4.4.3.1.

10.Install the sensor in the sample system and measure. If for some reason the Cal Adj value must be re-entered or modified to improve accuracy the unit does not have to be SpanChecked again.

55

Image 62
Contents Yyyyyyy y Sensor Serial No Instrument Serial No XDT User’s Manual Iii XDT User’s Manual Table of Contents Index Overview of the XDT XDT User’s Manual Sensor and Sampling Techniques Precautions using the sensor Sensor Technical Specifications Sensor Installation & Sampling TechniquesIn-situ Installation Not RecommendedExtractive Installation Mechanical Installation Troubleshooting unexpected readings SATInstrument Precautions using the XDT Electromagnetic Compatibility ConsiderationsInstrument Technical Specifications IsolationXDT-PM DIN43700 Enclosure Installation InstallationInstrument Mechanical Installation XDT-NEMA IP65 Enclosure InstallationConnecting Power Electrical ConnectionsWiring the Alarm Contacts AC Mains Electrical Power ConnectionLow Voltage DC Powered Option Electrical Power Connection Sensor ConnectionInterfacing to the Analog Output Starting up Operating the InstrumentInterfacing to the RS-232 option Push Buttons Display Conventions12 3 4 5 6 7 8 Viewing Dewpoint Mode Operating StateXDT User’s Manual Alarms HI **** ˚C Start CalibrationSpanCheck Mode Single Point Self Calibration, manual or scheduled Instrument XDT User’s Manual Viewing Serial Number Mode SetUp StateXDT User’s Manual Secured Password Protection Resetable Audio-Visual Alarm Option Nfpa compliant Troubleshooting the Instrument Display Explanation Required Action Maintenance XDT User’s Manual Glossary DB9 Glossary XDT User’s Manual Appendix a Flow Diagram of Operating State User Interface Sign on message Calibration without key pressHere Appendix B Flow Diagram of Set-Up State User Interface 16.24˚C˚FAppendix Sensor MechanicalAppendix D Optional Sensor Fittings Appendix E XDT Circuit Board Dimensions Side ViewAppendix E XDT-PM Enclosure Mounting Dimensions Appendix E XDT-NEMA Enclosure Mounting Dimensions Appendix F Sensor/SpanCheck Theory of Operation Appendix G Dewpoint Response time Analysis XDT User’s Manual Appendices Appendix H Sample Gas Filter Considerations Tested Instrument ˚CAppendix I XDT Grounding Considerations SensorAppendix J Analog Output vs. Dewpoint Instrument reading in ppmV Appendix K RS-232C Interface Protocol Help commandReceive Help Data Who Reset Login Data command Error SensOpen or Error SensShort or Error SensSat K n mWho Command ? commandSerial Number Command Login CommandAppendix L Procedure for Exchanging XDT Sensors State User InterfaceAppendix M Uncertainty in LBS & ppmV calculations Xentaur Return Authorization Number Request Form To obtain a RAN fill out this form and fax to 631Page Index Enclosure dimensions Pounds of H2O per million standard cubic Voltage, mains supply Iii Warranty