XDT User’s Manual Appendices
Appendix L: Procedure for Exchanging XDT Sensors
To exchange sensors one must know the measured low attenuation of the sensor at some low dew- point as well as the calibration adjustment value. These numbers can be found on the shipping container serial number label.
Through the following steps it is helpful to refer to Appendix A: Flow Diagram of XDT
State User Interface
1.Power up (or reset) the XDT while holding down the MODE key, in order to enter the
2.Press the MODE key to go to the “Choose Sensor Type” mode. Make sure that the proper sensor type is chosen, e.g. XT1 (for
3.Press the MODE key to go to the “Set Measured Attenuation @ low dewpoint” mode. The display will show the currently entered attenuation in decibels (dB), this is indicated by hav- ing both the °C and °F on the display. Use the UP or DOWN keys to select the value which appears on the shipping container label. Note that this value is a negative number typically in the range of
4.Press the MODE key to go to the “Set low dewpoint” mode. The display will show the cur- rently entered low dewpoint. Use the UP or DOWN keys to select the value which appears on the shipping container label.
5.Press the MODE key to go to the “Calibration Midrange Adjust” mode. The display will alternately show CAL and ADJ. Press the UP key once to see the entered calibration adjustment value. The display will alternately show the currently entered value and the dewpoint being measured, when showing the Cal Adj value the number (in the range of
6.Press the MODE key five (5) times to go to the “Lock or Unlock” mode. The display will alternately show UN and LOC. Press the DOWN key to lock the instrument. The display will alternately show only LOC.This prevents inadvertent changes.
7.Press the MODE key once.
8.Now the instrument has stored in it’s EEPROM, the calibration values associated with the new sensor, HOWEVER NOTE THAT IT IS STILL NOT CALIBRATED. The instrument may now be reset.
9.Perform the SpanCheck Procedure in section 3.4.4.3.1.
10.Install the sensor in the sample system and measure. If for some reason the Cal Adj value must be
55