Appendix H: Sample Gas Filter Considerations

Sample treatment with pre-filtering may be a necessity in many instances. However the filter housing and element may affect the dewpoint reading and/or the response time of the system. Below is a graph comparing measurements from the same calibrated sensor with and without a filter utilizing a polycarbonate bowl. All other parameters, which may affect the measurement, were kept constant. Outside of the sample system, a rather hot and humid environment was cre- ated, to illustrate a worse case condition. As can be seen the filter does not cause problems for dewpoints greater than -50°C. However the systems' ability to make precise measurements quickly deteriorates for dewpoints below -55°C.

Tested Instrument (˚C)

-35

Effect of polycarbonate filter on dewpoint measurement

Tests performed inside environmental chamber

 

-40

at 30˚C temperature and ~100%RH. Test gas

flow rate 3LPM. Using XDTsn: 2007 and

 

XTR-100 sensor sn: 18107

·Data with legend "ALL SS" was taken with an

-45

all stainless steel system.

 

· Data with legend "Through Filter" was taken with

-50

same system but with the addition of a polycarbonate

bowl filter

 

-55

 

-60

 

-65

 

-70

 

-75

MBW ˚C

 

-80

ALL SS ˚C

Through Filter ˚C

 

-85

 

-90

 

-90

-85 -80 -75 -70 -65 -60 -55 -50 -45 -40 -35

Chilled mirror reference instrument MBW (˚C)

With this in mind the user should select the appropriate filter construction materials. It is impor- tant to note that a stainless steel housing filter will not cause measurement errors, however if the filter element inside the housing is made of hygroscopic fibers, then there will be a delay intro- duced in the measurement.

48

Image 55
Contents Yyyyyyy y Sensor Serial No Instrument Serial No XDT User’s Manual Iii XDT User’s Manual Table of Contents Index Overview of the XDT XDT User’s Manual Sensor and Sampling Techniques Precautions using the sensor Sensor Installation & Sampling Techniques Sensor Technical SpecificationsNot Recommended In-situ InstallationExtractive Installation Mechanical Installation SAT Troubleshooting unexpected readingsElectromagnetic Compatibility Considerations Instrument Precautions using the XDTIsolation Instrument Technical SpecificationsXDT-NEMA IP65 Enclosure Installation InstallationInstrument Mechanical Installation XDT-PM DIN43700 Enclosure InstallationElectrical Connections Connecting PowerSensor Connection AC Mains Electrical Power ConnectionLow Voltage DC Powered Option Electrical Power Connection Wiring the Alarm ContactsInterfacing to the Analog Output Interfacing to the RS-232 option Operating the InstrumentStarting up 12 3 4 5 6 7 8 Display ConventionsPush Buttons Operating State Viewing Dewpoint ModeXDT User’s Manual Alarms Start Calibration HI **** ˚CSpanCheck Mode Single Point Self Calibration, manual or scheduled Instrument XDT User’s Manual SetUp State Viewing Serial Number ModeXDT User’s Manual Secured Password Protection Resetable Audio-Visual Alarm Option Nfpa compliant Troubleshooting the Instrument Display Explanation Required Action Maintenance XDT User’s Manual Glossary DB9 Glossary XDT User’s Manual Sign on message Calibration without key press Appendix a Flow Diagram of Operating State User InterfaceHere 16.24˚C˚F Appendix B Flow Diagram of Set-Up State User InterfaceSensor Mechanical AppendixAppendix D Optional Sensor Fittings Side View Appendix E XDT Circuit Board DimensionsAppendix E XDT-PM Enclosure Mounting Dimensions Appendix E XDT-NEMA Enclosure Mounting Dimensions Appendix F Sensor/SpanCheck Theory of Operation Appendix G Dewpoint Response time Analysis XDT User’s Manual Appendices Tested Instrument ˚C Appendix H Sample Gas Filter ConsiderationsSensor Appendix I XDT Grounding ConsiderationsAppendix J Analog Output vs. Dewpoint Instrument reading in ppmV Receive Help Data Who Reset Login Data command Help commandAppendix K RS-232C Interface Protocol K n m Error SensOpen or Error SensShort or Error SensSatLogin Command ? commandSerial Number Command Who CommandState User Interface Appendix L Procedure for Exchanging XDT SensorsAppendix M Uncertainty in LBS & ppmV calculations To obtain a RAN fill out this form and fax to 631 Xentaur Return Authorization Number Request FormPage Index Enclosure dimensions Pounds of H2O per million standard cubic Voltage, mains supply Iii Warranty