XDT User’s Manual Appendices
Appendix H: Sample Gas Filter Considerations
Sample treatment with
Tested Instrument (˚C)
Effect of polycarbonate filter on dewpoint measurement | ||
Tests performed inside environmental chamber | ||
| ||
at 30˚C temperature and ~100%RH. Test gas | ||
flow rate 3LPM. Using XDTsn: 2007 and | ||
|
·Data with legend "ALL SS" was taken with an
all stainless steel system. | ||
| · Data with legend "Through Filter" was taken with | |
same system but with the addition of a polycarbonate | ||
bowl filter | ||
| ||
| ||
| ||
| ||
| ||
MBW ˚C | ||
| ||
ALL SS ˚C | ||
Through Filter ˚C | ||
| ||
| ||
| ||
Chilled mirror reference instrument MBW (˚C)
With this in mind the user should select the appropriate filter construction materials. It is impor- tant to note that a stainless steel housing filter will not cause measurement errors, however if the filter element inside the housing is made of hygroscopic fibers, then there will be a delay intro- duced in the measurement.