Epson manual Instrument Mechanical Installation, XDT-PM DIN43700 Enclosure Installation

Page 18

Section 3: Instrument

3.3 Installation

3.3.1 Instrument Mechanical Installation

The XDT is available in several different physical configurations. Please follow the instructions below, which describe the instrument being installed. The sensor installation is discussed in Sec- tion 2: Sensor and Sampling Techniques.

3.3.1.1 XDT-PM (DIN43700) Enclosure Installation

The standard XDT-PM is provided with two clip-on compression panel mounting brackets. These brackets allow the user to mount and secure the instrument onto a properly cutout panel (137mm x 67mm). Consult with drawing in appendix E, for all relevant dimensions. Installation is accom- plished by attaching 2 clips, one on each side of instrument, and inserting the instrument into the panel cutout. Pressure must be applied along edges of instrument until it is firmly seated.

The XDT-PM may be ordered with an environmental seal option. This option provides a gasket (installed at the factory) to seal the face plate to the enclosure box, and a gasket (to be installed by the user) to seal the enclosure box to the user’s mounting panel. To achieve a good environmental seal the XDT-PM must be mounted onto a rigid flat panel with a cutout according to the specified dimensions, using the provided panel gasket and all four provided mounting brackets. The instructions listed below should be followed.

If the clip-on mounting brackets are installed on the instrument enclosure, remove them.

Make sure that the panel gasket is flat and even around the box bezel. Handle the gasket care- fully as not to tear it.

Insert the box into the user’s panel cutout, consult with drawing in appendix E for proper panel cutout dimensions.

Attach all four mounting brackets to the instrument enclosure from the rear of the panel, make sure they are fully locked in.

Use a screwdriver to tighten the mounting bracket screws, all four screws should be tightened uniformly, the panel gasket should be compressed.

3.3.1.2 XDT-NEMA (IP65) Enclosure Installation

The instrument can be installed as a wall or panel surface mount, by making use of four screw holes on the back (bottom) of the instrument enclosure. Open the instrument cover for access to these screw holes. They are located at the corners of the enclosure and outside of the NEMA seal. Do not drill other mounting holes, as you may compromise the seal. Refer to Appendix E for all relevant dimensions.

3.3.1.3 XDT-OEM bare board Installation

When mounting the circuit board in a user supplied enclosure using standoffs, be sure to use insu- lating washers. Refer to Appendix E for mounting hole dimensions and required mounting dimen- sions.

11

Image 18
Contents Yyyyyyy y Sensor Serial No Instrument Serial No XDT User’s Manual Iii XDT User’s Manual Table of Contents Index Overview of the XDT XDT User’s Manual Sensor and Sampling Techniques Precautions using the sensor Sensor Technical Specifications Sensor Installation & Sampling TechniquesIn-situ Installation Not RecommendedExtractive Installation Mechanical Installation Troubleshooting unexpected readings SATInstrument Precautions using the XDT Electromagnetic Compatibility ConsiderationsInstrument Technical Specifications IsolationXDT-PM DIN43700 Enclosure Installation InstallationInstrument Mechanical Installation XDT-NEMA IP65 Enclosure InstallationConnecting Power Electrical ConnectionsWiring the Alarm Contacts AC Mains Electrical Power ConnectionLow Voltage DC Powered Option Electrical Power Connection Sensor ConnectionInterfacing to the Analog Output Operating the Instrument Interfacing to the RS-232 optionStarting up Display Conventions 12 3 4 5 6 7 8Push Buttons Viewing Dewpoint Mode Operating StateXDT User’s Manual Alarms HI **** ˚C Start CalibrationSpanCheck Mode Single Point Self Calibration, manual or scheduled Instrument XDT User’s Manual Viewing Serial Number Mode SetUp StateXDT User’s Manual Secured Password Protection Resetable Audio-Visual Alarm Option Nfpa compliant Troubleshooting the Instrument Display Explanation Required Action Maintenance XDT User’s Manual Glossary DB9 Glossary XDT User’s Manual Appendix a Flow Diagram of Operating State User Interface Sign on message Calibration without key pressHere Appendix B Flow Diagram of Set-Up State User Interface 16.24˚C˚FAppendix Sensor MechanicalAppendix D Optional Sensor Fittings Appendix E XDT Circuit Board Dimensions Side ViewAppendix E XDT-PM Enclosure Mounting Dimensions Appendix E XDT-NEMA Enclosure Mounting Dimensions Appendix F Sensor/SpanCheck Theory of Operation Appendix G Dewpoint Response time Analysis XDT User’s Manual Appendices Appendix H Sample Gas Filter Considerations Tested Instrument ˚CAppendix I XDT Grounding Considerations SensorAppendix J Analog Output vs. Dewpoint Instrument reading in ppmV Help command Receive Help Data Who Reset Login Data commandAppendix K RS-232C Interface Protocol Error SensOpen or Error SensShort or Error SensSat K n mWho Command ? commandSerial Number Command Login CommandAppendix L Procedure for Exchanging XDT Sensors State User InterfaceAppendix M Uncertainty in LBS & ppmV calculations Xentaur Return Authorization Number Request Form To obtain a RAN fill out this form and fax to 631Page Index Enclosure dimensions Pounds of H2O per million standard cubic Voltage, mains supply Iii Warranty