Page 25

Site Preparation

Power requirements

Table 4 lists the power requirements for the 5975 Series MSD and related equipment. Extra power capacity for future additions is recommended.

Each product listed requires a dedicated circuit. The gauge controller is powered by the MSD. The MSD, GC, and data system must each have a separate circuit breaker. All of the equipment must share a common ground.

Power must meet the stability specifications listed in Table 4. Use a line monitor to check power stability. If your line power is unstable, you may need to install a line conditioner.

Table 4

Power requirements

 

 

 

 

 

 

 

 

 

Product

 

Line voltage

Maximum continuous

Supply

Outlets

 

 

 

power consumption

circuit rating

required

 

 

 

 

 

5975 Series MSD*

120-127 VAC 60 Hz

900 VA (400 VA for

15 A

1

 

 

 

foreline pump only)

 

 

 

 

200-240 VAC, 50/60 Hz

900 VA

15 A

1

 

 

 

 

 

7890A and 6890 GCs

Americas: 120 VACsingle phase

2250 VA

20 A

1

(standard oven) and ALS

(-10%/+5%), 48-66 Hz

 

 

 

 

 

220/230/240V single/split phase

2250 VA

15 A

1

 

 

(-10%/+5%), 48-66 Hz

 

 

 

 

 

 

 

 

7890A and 6890 GCs (fast

Japan: 200V split phase (-10%/+5%),

2950 VA

15 A

1

oven) and ALS

48-66 Hz

 

 

 

 

 

220/230/240 VACsingle/split phase

2950 VA

15 A

1

 

 

(-10%/+5%), 48-66 Hz

 

 

 

 

 

 

 

 

6850 GC (standard

America: 120 VAC, 48-63 Hz

1440 VA

15 A

1

oven)**

 

 

 

 

 

 

 

Japan: 100 VAC, 48-63 Hz

1440 VA

15 A

1

 

 

230 VAC, 48-63 Hz

2000 VA

8-12 A

1

 

 

 

 

 

6850 GC (fast oven)**

America: 120 VAC, 48-63 Hz

2400 VA

20 A

1

 

 

Japan: 200 VAC, 48-63 Hz

2400 VA

12 A

1

 

 

230 VAC, 48-63 Hz

2400 VA

11 A

1

 

 

 

 

 

 

Site Preparation Guide

25

Image 25
Contents Series Mass Selective Detectors U TI O N WA R N I N G Contents Power Cords General Information Instrument Identification Model number Product number Pump type Ionization modeMSD versions Many internal parts of the MSD carry dangerous voltages Important Safety WarningsMany parts are dangerously hot Electrostatic discharge is a threat to MSD electronicsWhen hydrogen gas is supplied to the instrument Hydrogen SafetyChemical Safety Information Safety and Regulatory CertificationsSymbols Sound emission declaration Electromagnetic compatibilityRecycling the Product CleaningSite Preparation Customer responsibility OverviewAgilent responsibility Compliance must exist prior to installationAnalysis Consulting services Other documentation Series MSD Troubleshooting and Maintenance ManualSeries MSD Series GC Computer with monitor Space and Weight RequirementsALS tray Product dimensions Telephone Communications RequirementsSite LAN network Voltage ranges of major components Electrical RequirementsConfiguration Measurement Nominal voltage Power configurationsSupply Outlets Power consumption Power requirementsProduct Line voltage RequiredSupply Outlets Power consumption Circuit rating Required Power plugs and cordOther electrical considerations Temperature Humidity relative Temperature, humidity, and altitudeAir Conditioning and Environmental Requirements Exhaust venting Airborne dustFume exhaust hood Purity Typical pressure range Typical flow, mL/min KPa psi Carrier and Reagent Gas RequirementsRegulators, tubing, and fittings Laboratory Supply Requirements Cleaning solventsData system supplies Part number Spare parts and consumablesMiscellaneous consumables Commonly used tools Chemical ConsumablesFerrules 5062-3506 Delivery and unloading Receiving the SystemInspecting for damage Unpacking StorageInstallation Installation and VerificationVerification Sensitivity specifications Power Cords Power Cords Page Agilent Technologies