Site Preparation

Regulators, tubing, and fittings

You must supply appropriate regulators for your carrier and reagent gas.

Regulators must be dual-stage.

Regulators must have stainless-steel (not elastomeric) diaphragms.

They must supply gas at the pressure specified.

They must have one outlet with 1/4-inch Swagelok fittings.

See the Agilent Technologies Chemical Analysis Columns and Supplies Catalog or visit the Agilent Technologies Chemical Analysis website at http://www.agilent.com/chem for dual-stage regulators available from Agilent Technologies.

You must supply Swagelok or Swagelok-compatible fittings, ferrules, and connectors for the 1/8-inch tubing.

32

Site Preparation Guide

Image 32
Contents Series Mass Selective Detectors U TI O N WA R N I N G Contents Power Cords General Information MSD versions Model number Product number Pump type Ionization modeInstrument Identification Important Safety Warnings Many internal parts of the MSD carry dangerous voltagesElectrostatic discharge is a threat to MSD electronics Many parts are dangerously hotHydrogen Safety When hydrogen gas is supplied to the instrumentChemical Safety Safety and Regulatory Certifications InformationSymbols Electromagnetic compatibility Sound emission declarationCleaning Recycling the ProductSite Preparation Overview Customer responsibilityCompliance must exist prior to installation Agilent responsibilityAnalysis Consulting services Series MSD Troubleshooting and Maintenance Manual Other documentationALS tray Space and Weight RequirementsSeries MSD Series GC Computer with monitor Product dimensions Site LAN network Communications RequirementsTelephone Electrical Requirements Voltage ranges of major componentsPower configurations Configuration Measurement Nominal voltagePower requirements Supply Outlets Power consumptionProduct Line voltage RequiredPower plugs and cord Supply Outlets Power consumption Circuit rating RequiredOther electrical considerations Air Conditioning and Environmental Requirements Temperature, humidity, and altitudeTemperature Humidity relative Airborne dust Exhaust ventingFume exhaust hood Carrier and Reagent Gas Requirements Purity Typical pressure range Typical flow, mL/min KPa psiRegulators, tubing, and fittings Data system supplies Cleaning solventsLaboratory Supply Requirements Miscellaneous consumables Spare parts and consumablesPart number Ferrules Chemical ConsumablesCommonly used tools 5062-3506 Inspecting for damage Receiving the SystemDelivery and unloading Storage UnpackingVerification Installation and VerificationInstallation Sensitivity specifications Power Cords Power Cords Page Agilent Technologies