Site Preparation

Space and Weight Requirements

Table 2 lists dimension and weight information for the MSD and related components. Your site must have enough bench space for the MSD, GC, data system, and accessories (Figure 1). In addition, there must be sufficient space around the system for ventilation and maintenance access. At least 30 cm

(12 inches) to the left of the MSD and at least 20 cm (8 1/4 inches) behind the MSD must be kept clear.

Benches must be sturdy enough to support the weight of the entire system.

The foreline pump can be located on the laboratory bench or on the floor. It must be close to the MSD because it is connected by a 200-cm (79-inch) hose. The hose is stiff and cannot be bent sharply.

C A U TI O N

Do not put the foreline pump on your laboratory bench if vibration-sensitive equipment

is located on the bench.

 

5975 Series MSD

6890 Series GC

Computer with monitor

Printer

ALS tray

Figure 1 Top view of typical installation

20

Site Preparation Guide

Image 20
Contents Series Mass Selective Detectors U TI O N WA R N I N G Contents Power Cords General Information MSD versions Model number Product number Pump type Ionization modeInstrument Identification Important Safety Warnings Many internal parts of the MSD carry dangerous voltagesElectrostatic discharge is a threat to MSD electronics Many parts are dangerously hotHydrogen Safety When hydrogen gas is supplied to the instrumentChemical Safety Safety and Regulatory Certifications InformationSymbols Electromagnetic compatibility Sound emission declarationCleaning Recycling the ProductSite Preparation Overview Customer responsibilityCompliance must exist prior to installation Agilent responsibilityAnalysis Consulting services Series MSD Troubleshooting and Maintenance Manual Other documentationALS tray Space and Weight RequirementsSeries MSD Series GC Computer with monitor Product dimensions Site LAN network Communications RequirementsTelephone Electrical Requirements Voltage ranges of major componentsPower configurations Configuration Measurement Nominal voltagePower requirements Supply Outlets Power consumptionProduct Line voltage RequiredPower plugs and cord Supply Outlets Power consumption Circuit rating RequiredOther electrical considerations Air Conditioning and Environmental Requirements Temperature, humidity, and altitudeTemperature Humidity relative Airborne dust Exhaust ventingFume exhaust hood Carrier and Reagent Gas Requirements Purity Typical pressure range Typical flow, mL/min KPa psiRegulators, tubing, and fittings Data system supplies Cleaning solventsLaboratory Supply Requirements Miscellaneous consumables Spare parts and consumablesPart number Ferrules Chemical ConsumablesCommonly used tools 5062-3506 Inspecting for damage Receiving the SystemDelivery and unloading Storage UnpackingVerification Installation and VerificationInstallation Sensitivity specifications Power Cords Power Cords Page Agilent Technologies