Agilent Technologies 5975 manual Other electrical considerations

Page 27

Site Preparation

WA R N I N G

Do not use extension cords with the MSD or GC. Extension cords normally are not

rated to carry enough power and can be a safety hazard.

 

 

 

N O T E

Maintain easy access to the power cords so they can be disconnected for maintenance.

Other electrical considerations

Additional electrical considerations include:

Electromagnetic interference (EMI), such as is generated by NMRs, radio transmitters, and microwave links, may interfere with system performance.

Protect the system from static electricity by observing humidity and temperature requirements. Minimize the presence of nonconductive products such as carpets and vinyl floor tiles.

Install emergency-off pushbuttons that can disconnect power to the ventilation system and all electric equipment in the room except overhead lighting.

Provide separate convenience outlets for building maintenance and other appliances. Convenience outlets must be on circuits separate from the MSD system. Convenience outlets must share the normal building distribution ground, not the MSD system ground.

In some geographical areas it may be advisable to install lightning protection for personnel and equipment.

Site Preparation Guide

27

Image 27
Contents Series Mass Selective Detectors U TI O N WA R N I N G Contents Power Cords General Information Model number Product number Pump type Ionization mode Instrument IdentificationMSD versions Many internal parts of the MSD carry dangerous voltages Important Safety WarningsMany parts are dangerously hot Electrostatic discharge is a threat to MSD electronicsWhen hydrogen gas is supplied to the instrument Hydrogen SafetyChemical Safety Information Safety and Regulatory CertificationsSymbols Sound emission declaration Electromagnetic compatibilityRecycling the Product CleaningSite Preparation Customer responsibility OverviewAgilent responsibility Compliance must exist prior to installationAnalysis Consulting services Other documentation Series MSD Troubleshooting and Maintenance ManualSpace and Weight Requirements Series MSD Series GC Computer with monitorALS tray Product dimensions Communications Requirements TelephoneSite LAN network Voltage ranges of major components Electrical RequirementsConfiguration Measurement Nominal voltage Power configurationsRequired Power requirementsSupply Outlets Power consumption Product Line voltageSupply Outlets Power consumption Circuit rating Required Power plugs and cordOther electrical considerations Temperature, humidity, and altitude Temperature Humidity relativeAir Conditioning and Environmental Requirements Exhaust venting Airborne dustFume exhaust hood Purity Typical pressure range Typical flow, mL/min KPa psi Carrier and Reagent Gas RequirementsRegulators, tubing, and fittings Cleaning solvents Laboratory Supply RequirementsData system supplies Spare parts and consumables Part numberMiscellaneous consumables Chemical Consumables Commonly used toolsFerrules 5062-3506 Receiving the System Delivery and unloadingInspecting for damage Unpacking StorageInstallation and Verification InstallationVerification Sensitivity specifications Power Cords Power Cords Page Agilent Technologies