Site Preparation |
|
|
|
| |
Table 4 | Power requirements (continued) |
|
|
| |
|
|
|
|
|
|
Product |
| Line voltage | Maximum continuous | Supply | Outlets |
|
|
| power consumption | circuit rating | required |
|
|
|
|
| |
ChemStation computer | 1000 VA | 15 A | 3 to 5 | ||
|
| 50/60 Hz ± 5% |
|
|
|
|
|
|
|
|
|
*The MSD operates on either voltage range. The foreline and turbo pumps are supplied for the voltage range suitable to the country where the order originates
† Americas 120 V requires 20 A dedicated line. Americas 240 V requires 15 A dedicated line.
‡Option 003, fast oven, is a unit with operating range 198 to 231 V. Most laboratories have
** Requires an isolated ground and dedicated outlet.
Power plugs and cord
| The MSD is supplied with a power cord and plug appropriate for the country | |
| from which the order originates. For example, if an order originates from a | |
| Agilent Technologies sales office in Germany, the power cord and plug | |
| supplied will be compatible with the standard voltage and outlet configuration | |
| in Germany. See “Power Cords" on page 41 for illustrations of the power cords | |
| available. | |
| Data system components also include power cords with plugs appropriate for | |
| the country where the order was placed. | |
| Power cord lengths for the MSD and the data system components and | |
| accessories are approximately 2.3 m (7.5 ft). | |
| If an instrument is being ordered from one location but is to be installed in another | |
C A U TI O N | ||
location with different electrical power characteristics, this must be noted on the order. | ||
| ||
| A special note must also be made if the electrical power at the site is different from the | |
| standard electrical power in that country. | |
|
| |
| Make sure the power cords supplied with the MSD are appropriate for your country | |
WA R N I N G | ||
and site before installing the instrument. See “Power Cords” on page 41. | ||
| ||
|
|
26 | Site Preparation Guide |