Agilent Technologies 5975 manual Power plugs and cord

Page 26

Site Preparation

 

 

 

 

Table 4

Power requirements (continued)

 

 

 

 

 

 

 

 

 

Product

 

Line voltage

Maximum continuous

Supply

Outlets

 

 

 

power consumption

circuit rating

required

 

 

 

 

 

ChemStation computer

100-120 or 220-240 VAC+5/-10%,

1000 VA

15 A

3 to 5

 

 

50/60 Hz ± 5%

 

 

 

 

 

 

 

 

 

*The MSD operates on either voltage range. The foreline and turbo pumps are supplied for the voltage range suitable to the country where the order originates

† Americas 120 V requires 20 A dedicated line. Americas 240 V requires 15 A dedicated line.

Option 003, fast oven, is a unit with operating range 198 to 231 V. Most laboratories have 4-wire service resulting in 208 V at the wall receptacle. It is important to measure the line voltage at the receptacle for the GC.

** Requires an isolated ground and dedicated outlet.

Power plugs and cord

 

The MSD is supplied with a power cord and plug appropriate for the country

 

from which the order originates. For example, if an order originates from a

 

Agilent Technologies sales office in Germany, the power cord and plug

 

supplied will be compatible with the standard voltage and outlet configuration

 

in Germany. See “Power Cords" on page 41 for illustrations of the power cords

 

available.

 

Data system components also include power cords with plugs appropriate for

 

the country where the order was placed.

 

Power cord lengths for the MSD and the data system components and

 

accessories are approximately 2.3 m (7.5 ft).

 

If an instrument is being ordered from one location but is to be installed in another

C A U TI O N

location with different electrical power characteristics, this must be noted on the order.

 

 

A special note must also be made if the electrical power at the site is different from the

 

standard electrical power in that country.

 

 

 

Make sure the power cords supplied with the MSD are appropriate for your country

WA R N I N G

and site before installing the instrument. See “Power Cords” on page 41.

 

 

 

26

Site Preparation Guide

Image 26
Contents Series Mass Selective Detectors U TI O N WA R N I N G Contents Power Cords General Information MSD versions Model number Product number Pump type Ionization modeInstrument Identification Important Safety Warnings Many internal parts of the MSD carry dangerous voltagesElectrostatic discharge is a threat to MSD electronics Many parts are dangerously hotHydrogen Safety When hydrogen gas is supplied to the instrumentChemical Safety Safety and Regulatory Certifications InformationSymbols Electromagnetic compatibility Sound emission declarationCleaning Recycling the ProductSite Preparation Overview Customer responsibilityCompliance must exist prior to installation Agilent responsibilityAnalysis Consulting services Series MSD Troubleshooting and Maintenance Manual Other documentationALS tray Space and Weight RequirementsSeries MSD Series GC Computer with monitor Product dimensions Site LAN network Communications RequirementsTelephone Electrical Requirements Voltage ranges of major componentsPower configurations Configuration Measurement Nominal voltageProduct Line voltage Power requirementsSupply Outlets Power consumption RequiredPower plugs and cord Supply Outlets Power consumption Circuit rating RequiredOther electrical considerations Air Conditioning and Environmental Requirements Temperature, humidity, and altitudeTemperature Humidity relative Airborne dust Exhaust ventingFume exhaust hood Carrier and Reagent Gas Requirements Purity Typical pressure range Typical flow, mL/min KPa psiRegulators, tubing, and fittings Data system supplies Cleaning solventsLaboratory Supply Requirements Miscellaneous consumables Spare parts and consumablesPart number Ferrules Chemical ConsumablesCommonly used tools 5062-3506 Inspecting for damage Receiving the SystemDelivery and unloading Storage UnpackingVerification Installation and VerificationInstallation Sensitivity specifications Power Cords Power Cords Page Agilent Technologies