Agilent Technologies G1978B setup guide Quad Performance Verification Summary, Scan Mode

Page 27

Verification 2

To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes

To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes

Preparation

1Transfer 1 mL of 5 ng/µL reserpine (Agilent G2423A) to a 50 mL volumetric flask. Use a clean graduated pipette.

2Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium formate.

3Transfer 5 mL of the first dilution to a 50 mL volumetric flask. Use a clean graduated pipette.

4Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium formate.

This provides the final 10 pg/µL reserpine concentration required for performance verification.

5Transfer approximately 1 mL of the second dilution to an autosampler vial.

Table 3

6130 and 6140 Quad Performance Verification Summary, Scan Mode

 

 

 

 

 

 

G1978B

G1978B

 

 

ES Positive Scan Mode

APCI Positive Scan Mode

 

 

 

 

Sample

 

Reserpine, 5 ng/µL

Reserpine, 5 ng/µL

 

 

 

Concentration after dilution

10 pg/µL

10 pg/µL

 

 

 

Injection volume

5 µL

5 µL

 

 

 

Total sample amount injected

50 pg

50 pg

 

 

 

Sample order number

G2423A

G2423A

 

 

 

 

Solvent

 

75:25

75:25

 

 

methanol/water

methanol/water

 

 

with 5mM

with 5mM

 

 

ammonium formate

ammonium formate

 

 

 

 

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

27

Image 27
Contents Set-Up Guide U T I O N WA R N I N G This Guide Page Content Contents Installation Prepare to install InstallationFrom left to right G1946-80058, G1960-60802 and G1960-60856 Install the HV control PCA and cablesConnecting the valve PCA to the multimode HV power supply Connecting the Vcap and Vchamber cables to the power supply Connecting the HV PCA to the chassis To remove the multimode source Changing SourcesTo convert from multimode to ESI, Apci or Appi To convert from ESI, Apci or Appi to the multimode source Installation onStandard spray shield and capillary cap for ESI or Apci T E Multimode Spray Chamber No nebulizer on top of the multimode source LC/MS sample tubing connected to LC/MS inlet filter Verification Preparation 6110 and 6120 Quad Performance Verification Summary Table 6140 Quad Performance Verification Summary, SIM Mode 20 1 pk-pk 10 1 pk-pk Verification 6140 Quad Performance Verification Summary, Scan Mode 6130SCNESMM.M 6130SCNCIMM.M Verification Create and run a Multiple FIA Method Sequence Verification Autotune report To do an autotuneExample of multimode verification report Example of multimode verification report Methods Be impaired and the system may catch fire To setup a method to use the multimode sourceMethods Polarity options for Signal 1 and Signal Method Spray Chamber set to MM-ES+APCI To create a method for alternating ESI and Apci operationIonization mode for Signal 1 and Signal Index Index Page G1978-90070