Agilent Technologies G1978B setup guide Connecting the HV PCA to the chassis

Page 12

1Installation

Step 2. Install the HV control PCA and cables

Figure 9 Connecting the HV PCA to the chassis.

12 Close the AUX Module cover and reconnect all cables.

13 Install the multimode source to the system and connect all connectors.

Figure 10 Installing the multimode source (left) and connecting all connectors.

14 Put back the side, top, front and CDS cover.

15 Plug the system power cord back on and turn the front switch on.

The pump down process will start.

16 Start the ChemStation program.

17 Click Method and Run Control View > MSD > Spray Chamber and verify that the source is an MM Source.

18 Go to MSD Tune window and start the Auto tune with a multimode source. 19 Verify that the multimode source is working correctly.

12

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

Image 12
Contents Set-Up Guide U T I O N WA R N I N G This Guide Page Content Contents Installation Installation Prepare to install Install the HV control PCA and cables From left to right G1946-80058, G1960-60802 and G1960-60856Connecting the valve PCA to the multimode HV power supply Connecting the Vcap and Vchamber cables to the power supply Connecting the HV PCA to the chassis Changing Sources To remove the multimode sourceTo convert from multimode to ESI, Apci or Appi Installation on To convert from ESI, Apci or Appi to the multimode sourceStandard spray shield and capillary cap for ESI or Apci T E Multimode Spray Chamber No nebulizer on top of the multimode source LC/MS sample tubing connected to LC/MS inlet filter Verification Preparation 6110 and 6120 Quad Performance Verification Summary Table 6140 Quad Performance Verification Summary, SIM Mode 20 1 pk-pk 10 1 pk-pk Verification 6140 Quad Performance Verification Summary, Scan Mode 6130SCNESMM.M 6130SCNCIMM.M Verification Create and run a Multiple FIA Method Sequence Verification To do an autotune Autotune reportExample of multimode verification report Example of multimode verification report Methods To setup a method to use the multimode source Be impaired and the system may catch fireMethods Polarity options for Signal 1 and Signal To create a method for alternating ESI and Apci operation Method Spray Chamber set to MM-ES+APCIIonization mode for Signal 1 and Signal Index Index Page G1978-90070