Agilent Technologies G1978B setup guide Methods

Page 37

Methods 3

To create a method for positive/negative mixed mode operation

To create a method for positive/negative mixed mode operation

1Open the MSD Spray Chamber dialog box by clicking the Instrument > Set Up MSD Signals in the Method and Run Control view.

2Select MM-ES_APCI from the Method Spray Chamber drop-down list.

3Check that Installed Spray Chamber is also set to MM-ES+APCI.

4Make any other changes that are necessary for your method.

5Click the OK button.

6Open the Set up MSD Signals dialog box by clicking the Instrument > MSD Spray chamber in the Method and Run Control view.

7Modify the settings so that Signal 1 has Positive polarity and Signal 2 has Negative polarity as shown in Figure 21.

8Make any other changes that are necessary for your method.

9Click the OK button.

Fast positive/negative polarity switching is a very useful technique but it requires time for the ion chemistry to be established and the optics path to refill with ions. The gas density plays a role in the speed of refilling the ion path. The gas density is affected by source temperature. For a method running positive/negative switching, use a lower vaporizer temperature (150 to 200°C) and a lower Vcap (approximately 1000 V). These will greatly affect the quality of the results in positive/negative switching experiments.

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

37

Image 37
Contents Set-Up Guide U T I O N WA R N I N G This Guide Page Content Contents Installation Prepare to install InstallationFrom left to right G1946-80058, G1960-60802 and G1960-60856 Install the HV control PCA and cablesConnecting the valve PCA to the multimode HV power supply Connecting the Vcap and Vchamber cables to the power supply Connecting the HV PCA to the chassis To remove the multimode source Changing SourcesTo convert from multimode to ESI, Apci or Appi To convert from ESI, Apci or Appi to the multimode source Installation onStandard spray shield and capillary cap for ESI or Apci T E Multimode Spray Chamber No nebulizer on top of the multimode source LC/MS sample tubing connected to LC/MS inlet filter Verification Preparation 6110 and 6120 Quad Performance Verification Summary Table 6140 Quad Performance Verification Summary, SIM Mode 20 1 pk-pk 10 1 pk-pk Verification 6140 Quad Performance Verification Summary, Scan Mode 6130SCNESMM.M 6130SCNCIMM.M Verification Create and run a Multiple FIA Method Sequence Verification Autotune report To do an autotuneExample of multimode verification report Example of multimode verification report Methods Be impaired and the system may catch fire To setup a method to use the multimode sourceMethods Polarity options for Signal 1 and Signal Method Spray Chamber set to MM-ES+APCI To create a method for alternating ESI and Apci operationIonization mode for Signal 1 and Signal Index Index Page G1978-90070