Agilent Technologies G1978B setup guide Verification

Page 29

Verification 2

To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes

When the method is finished, a report prints showing the signal-to-noise ratio for the five blank and five sample peaks, and a blank-subtracted average of the sample peaks. The blank-subtracted average signal-to-noise value must be greater than or equal to 20:1 peak-to-peak (100:1 RMS).

Sensitivity Verification for Multimode in APCI, Positive Scan Mode

1Start the ChemStation software.

2Change the view to MSD Tune, and start an Autotune.

After the autotune completes, you may need to wait up to 30 minutes to allow for the calibrant solution to be pumped out of the Quad instrument. This minimizes any background signal resulting from the calibrant.

3Change the view to Method and Run Control, and load the method 6130SCNCI_MM.M (6130 model) or 6140SCNCI_MM.M (6140 model).

4Edit the method to ensure that 75:25 methanol/water with 5mM ammonium formate is selected as the LC solvent.

5Place the vials into the LC autosampler.

Position #1: empty, uncapped vial

Position #2: vial of the solvent used for dilution (solvent blank)

Position #3: vial with the reserpine sample (10 pg/µL)

6Run the method.

The method does an FIA run with one injection of the empty vial, five injections of the solvent blank, and five injections of the reserpine sample.

You may need to further optimize the nebulizer pressure by running FIA experiments to achieve maximum instrument sensitivity

7Review the results.

When the method is finished, a report prints showing the signal-to-noise ratio for the five blank and five sample peaks, and a blank-subtracted average of the sample peaks. The blank-subtracted average signal-to-noise value must be greater than or equal to 10:1 peak-to-peak (50:1 RMS).

Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide

29

Image 29
Contents Set-Up Guide U T I O N WA R N I N G This Guide Page Content Contents Installation Prepare to install InstallationFrom left to right G1946-80058, G1960-60802 and G1960-60856 Install the HV control PCA and cablesConnecting the valve PCA to the multimode HV power supply Connecting the Vcap and Vchamber cables to the power supply Connecting the HV PCA to the chassis To remove the multimode source Changing SourcesTo convert from multimode to ESI, Apci or Appi To convert from ESI, Apci or Appi to the multimode source Installation onStandard spray shield and capillary cap for ESI or Apci T E Multimode Spray Chamber No nebulizer on top of the multimode source LC/MS sample tubing connected to LC/MS inlet filter Verification Preparation 6110 and 6120 Quad Performance Verification Summary Table 6140 Quad Performance Verification Summary, SIM Mode 20 1 pk-pk 10 1 pk-pk Verification 6140 Quad Performance Verification Summary, Scan Mode 6130SCNESMM.M 6130SCNCIMM.M Verification Create and run a Multiple FIA Method Sequence Verification Autotune report To do an autotuneExample of multimode verification report Example of multimode verification report Methods Be impaired and the system may catch fire To setup a method to use the multimode sourceMethods Polarity options for Signal 1 and Signal Method Spray Chamber set to MM-ES+APCI To create a method for alternating ESI and Apci operationIonization mode for Signal 1 and Signal Index Index Page G1978-90070