Output
Current
Monitor
Circuit
Calibration
Calibration
Calibrating for Readback Accuracy
1.Ensure that the power supply is turned OFF. Disconnect any load.
2.Set SW1 switch 6 OPEN to select
3.Connect the shunt and DVM across the power supply output to read the output current. Connect a second DVM between the rear panel J2 connector terminals 12 (output current monitor) and 11 (current monitor signal return).
4.Turn the current control fully counter clockwise, then turn on the power supply.
5.Set the power supply’s output current to 1% of the
6.Adjust the output current monitor offset (R394) until the monitor DVM indicates 1% of the readback scale. (50 mVdc for
7.Set the power supply output current to the
8.Adjust the output current monitor range (R383) until the monitor DVM indicates 100% of the readback scale. (5 Vdc for
Repeat this procedure until the monitor signal is adjusted to power supply specifications (see Section 1) or to the accuracy that your application requires.
Note:Calculate the required DVM reading using V= IR where V is the DVM reading, I is the current, and R is the DC shunt resistance.
Release 2.1 | 85 |