Omega CL526 manual Battery charger. Operation from line source, Resistance and Rtd measurements

Page 16

P

Keyboard

D/A

1000 mV

x 1000 multipl.

100 mV

54 mV

22 mV

Out

4.8Battery charger. Operation from line source

ATTENTION: ONLY FOR USE WITH NI-MH BATTERIES.

The auxiliary module, supplied as a standard accessory, allows operation from 110-120 Vac or 220-240 Vac 50/60 Hz. The calibrator, if needed, can be operated directly from a line source through the charger. The plastic case of the battery charger incorporates the line voltage plug and a cable with connector for interconnection to the instrument. The charger circuit is designed with an insulating transformer and a voltage stabilizer circuit.

The step-down transformer reduces the power line (110-120 Vac or 220-240 Vac nominal) to a value of 10 Vac. The above voltage is full wave rectified, filtered and stabilized. The output voltage of 6,6 V is the ideal value to recharge the internal Ni-MH batteries.

4.9Digital interface

The digital interface circuit is essentially based on the serial communication interface subsystem (SCI) on the chip of the microprocessor at 0 to +5V level.

An adaptor to convert TTL to RS 232 voltage levels can be obtained on request.

4.10Resistance and Rtd measurements

The resistance thermometer (Rtd) is connected to terminals A-B-C in a 3-wire configuration (see figure on the next page).

Two constant current generators are provided by the auxiliary module for supplying the Rtd.

The first half of “IC1” generates the negative current IA= - 0.25 mA that flows from terminal B to terminal A through the Rtd and line resistances RLA and RLB. IA is kept constant by the microprocessor that controls the zero voltage level. The second half of “IC 1”, with the associated resistors, generates the positive current IC that flows from terminal C to terminal B through line resistances RLC and RLB. Current IC is kept exactly = 2 x IA, so the resultant current IB = IC - IA flows through RLB. The input measured signal across terminals A and B is the algebraic sum of drop voltages across Rtd and line resistances RLA and RLB. As drop voltages across RLA and RLB are exactly the same (providing that line resistances RLA and RLB are equals), but with opposite poles, the resultant voltage across terminals A and B is proportional to Rtd resistance variation, with no influence of line resistance.

The measured signal is then handled by the microprocessor that linearizes it and displays the corresponding value in engineering units.

 

 

IA = - 0.25 mA

A

 

IC1/1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

IB = IC - IA = + 0.25 mA

B

 

 

 

 

 

 

 

 

 

 

IC = + 0.5 mA

C

 

 

IC1/2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RLA

Rtd

RLB

RLC

16

Image 16
Contents CL526 OMEGAnetSM On-Line Service Internet e-mail Info@omega.com Introductory Note Table of Contents Digital Interface General Performance Specifications Power supply Battery lifeBattery voltage Dimensions 120x60x230 mm Weights net 1 KgTable of ranges and accuracy 32 to +302F ± 0.02% of rdg +0.18FGeneral Features Case Physical Description Functional Description Power supply+ 5 V for logic and analog circuits 5 V for analog circuits Input circuit Microprocessor FirmwareDigital display Digital to analog converterBattery charger. Operation from line source Resistance and Rtd measurementsResistance and Rtd simulation Thermocouples input-output circuitUnpacking PRE-OPERATIONAL Check Out 1248.3C TcWiring practice Electrical ConnectionsThermocouple wires Examples of input /output wiring and connectionsLimit of Error of thermocouple wires ClassPower Supply Battery ChargerHow to maximize the battery life Battery Line OpPower on Battery voltage indicationOperating mode set up Operation & ApplicationsOUT mode selection Parameter or sensor selection1032.2C TcK Out 0.0C TcKInternational Temperature Scale Rj modeTecnical unit Decimal point positionAverage readings Data memory configurationIN-OUT data memories Data memory manual recall Data memory automatic scanningAutomatic simulation cycle Manual step advanceSimulation cycle selection Time 0h 0m 50s Start 0.0mVEnd 100.0mV 0mVRj compensation mode check Simulation cycleMode Linear Mode Square Scale factor programRJ 0.0C ext ProgWord mbar Word psi Word hPaFirmware version code Serial number Installation parameter procedureExternal Rj compensation Digital interface data program mode Digital output wiring practiceDigital Interface Baud RateCommunication protocol from CL526 to a PC TTL to RS 232 adapterComputer CL526 Instruct DataPage Ramp Computer request for CL526 settings = OHMInstr ModeDis/Lin STO 2 #D 127 202 Display Chksum = DATA1 + DATA2 + Data 3 + Data 4 .AND.7F Communication programs Example a data transfer from CL526 to PCExample B PC instructions to CL526 Example a3ØØ While LOC 1 = Ø Wend Maintenance Safety recommendationsFaulty operating conditions Error ChecksumProtection fuses StorageIndex WARRANTY/DISCLAIMER Temperature