Omega CL526 manual Time 0h 0m 50s, Start 0.0mV, End 100.0mV, Soak 0h 0m 50s

Page 30

End

 

 

 

 

 

 

Z step

Start

 

 

 

Time

Soak

Time

Soak

To memory load the cycle parameters, follow the procedure indicated below

Select the required tecnical unit or electrical parameters

Select the required decimal point position

Press <SHIFT> + <TIME> keys to enter the cycle set-up procedure obtaining the following indication

Time 0h 0m 50s

related to the "time" in hours, minutes and seconds. The maximum setting is limited to 5 hours 33 minutes 20 seconds (20.000 seconds)

Press the <START> key to obtain the following indication:

Start 0.0mV

Press the <▲> and <▼> cursors to set the "Start" level of the cycle

Press <ENTER>+<START> keys to memory store the new value

Press the <END> key to obtain the following indication:

End 100.0mV

Press <▲> and <▼> cursors to set the "end" level of the cycle

Press <ENTER>+<END> keys to memory store the new value

Press the <STEP> key to obtain the following indication:

Step1.0mV

Press <▲> and <▼> cursors to set the size of each individual step. To have a continuous ramp set the minimum possible value (e.g. = 0.1)

Press <ENTER>+<STEP> keys to memory store the new value

Press the <TIME> key to obtain the following indication:

Time 0h 0m 50s

Press <▲> and <▼> cursors to set the required ramp time - max. 5h-33m-20s (20000 seconds)

Press <ENTER>+<TIME> keys to memory store the new value

A setting of 0h-0m-00s allows a manual step advance each time the <AUTORAMP> key is pressed.

Press the <SOAK> key to read the following indication:

Soak 0h 0m 50s

Press <▲> and <▼> cursors to set the waiting time (or soak time). If the setting is 0h-0m-0s the waiting or soak time is excluded

30

Image 30
Contents CL526 OMEGAnetSM On-Line Service Internet e-mail Info@omega.com Introductory Note Table of Contents Digital Interface General Performance Specifications Battery voltage Power supplyBattery life Dimensions 120x60x230 mm Weights net 1 KgTable of ranges and accuracy 32 to +302F ± 0.02% of rdg +0.18FGeneral Features Case Physical Description Power supply Functional Description+ 5 V for logic and analog circuits 5 V for analog circuits Input circuit Digital display MicroprocessorFirmware Digital to analog converterBattery charger. Operation from line source Resistance and Rtd measurementsResistance and Rtd simulation Thermocouples input-output circuitUnpacking PRE-OPERATIONAL Check Out 1248.3C TcWiring practice Electrical ConnectionsThermocouple wires Examples of input /output wiring and connectionsLimit of Error of thermocouple wires ClassHow to maximize the battery life Power SupplyBattery Charger Battery Line OpOperating mode set up Power onBattery voltage indication Operation & Applications1032.2C TcK OUT mode selectionParameter or sensor selection Out 0.0C TcKTecnical unit International Temperature ScaleRj mode Decimal point positionData memory configuration Average readingsIN-OUT data memories Data memory manual recall Data memory automatic scanningManual step advance Automatic simulation cycleSimulation cycle selection End 100.0mV Time 0h 0m 50sStart 0.0mV 0mVRj compensation mode check Simulation cycleRJ 0.0C ext Mode Linear Mode SquareScale factor program ProgWord mbar Word psi Word hPaInstallation parameter procedure Firmware version code Serial numberExternal Rj compensation Digital Interface Digital interface data program modeDigital output wiring practice Baud RateTTL to RS 232 adapter Communication protocol from CL526 to a PCComputer CL526 Instruct DataPage Ramp Computer request for CL526 settings = OHMInstr ModeDis/Lin STO 2 #D 127 202 Display Chksum = DATA1 + DATA2 + Data 3 + Data 4 .AND.7F Example B PC instructions to CL526 Communication programsExample a data transfer from CL526 to PC Example a3ØØ While LOC 1 = Ø Wend Faulty operating conditions MaintenanceSafety recommendations Error ChecksumProtection fuses StorageIndex WARRANTY/DISCLAIMER Temperature