1.2.1Table of ranges and accuracy
Sensor or | Total range | High accuracy | Resolution | Limit of error |
parameter |
| range |
|
|
Tc type J | 0.1°C | ± (0.02% of rdg +0.1°C) | ||
| 0.1°F | ± (0.02% of rdg +0.18°F) | ||
Tc type K | 0.1°C | ± (0.02% of rdg +0.1°C) | ||
| 0.1°F | ±(0.02% of rdg +0.18°F) | ||
Tc type T | 0.1°C | ± (0.02% of rdg +0.1°C) | ||
| 0.1°F | ±0.02% of rdg +0.18°F) | ||
Tc type R | 0 to +1760°C | 500 to 1700°C | 0.1°C | ± (0.02% of rdg +0.3°C) |
| 32 to +3200°F | 932 to 3092°F | 0.1°F | ± (0.02% of rdg +0.154°F) |
Tc type S | 0 to +1760°C | 600 to 1760°C | 0.1°C | ± (0.02% of rdg +0.3°C) |
| 32 to +3200°F | 1112 to 3200°F | 0.1°F | ± (0.02% of rdg +0.154°F) |
Tc type B | 200 to +1820°C | 1000 to 1820°C | 0.1°C | ±(0.02% of rdg +0.4°C) |
| 392 to +3308°F | 1832 to 3308°F | 0.1°F | ± (0.02% of rdg +0.172°F) |
Tc type C | 0 to +2300°C | 1150 to 2300°C | 0.1°C | ± (0.02% of rdg +0.4°C) |
| 32 to +4172°F | 2102 to 4172°F | 0.1°F | ± (0.02% of rdg +0.172°F) |
Tc type G | 0 to +2300°C | 300 to 2000°C | 0.1°C | ± (0.02% of rdg +0.4°C) |
| 32 to +4172°F | 572 to 3632°F | 1°F | ± (0.02% of rdg +0.172°F) |
Tc type D | 0 to +2300°C | 300 to 2000°C | 0.1°C | ± (0.02% of rdg +0.4°C) |
| 32 to +4172°F | 572 to 3632°F | 0.1°F | ± (0.02% of rdg +0..72°F) |
Tc type U | 0.1°C | ±(0.02% of rdg +0.1°C) | ||
| 0.1°F | ± (0.02% of rdg +0.18°F) | ||
Tc type L | 0.1°C | ± (0.02% of rdg +0.1°C) | ||
| 0.1°F | ± (0.02% of rdg +0.18°F) | ||
Tc type N | 0 to 1300°C | 0 to 1300°C | 0.1°C | ±(0.02% of rdg +0.1°C) |
| +32 to 2372°C | 32 to 2372 | 0.1°F | ± (0.02% of rdg +0.18°F) |
Tc type E | 0.1°C | ±(0.02% of rdg +0.1°C) | ||
| 0.1°F | ± (0.02% of rdg +0.18°F) | ||
Tc type F | 0 to 1400°C | 0 to 1400°C | 0.1°C | ±(0.02% of rdg +0.1°C) |
| 32 to 2552°F | 32 to 2552°F | 0.1°F | ± (0.02% of rdg +0.18°F) |
Pt 100 (IEC) | 0.1°C | ±(0.02% of rdg +0.1°C) | ||
| 0.1°F | ± (0.02% of rdg +0.18°F) | ||
Pt100 (JIS) | 0.1°C | ±(0.02% of rdg +0.1°C) | ||
| 0.1°F | ± (0.02% of rdg +0.18°F) | ||
Pt100 (US) | 0.1°C | ±(0.02% of rdg +0.1°C) | ||
| 0.1°F | ± (0.02% of rdg +0.18°F) | ||
Ni100 | 0.1°C | ±(0.02% of rdg +0.1°C) | ||
|
| 0.1°F | ± (0.02% of rdg +0.18°F) | |
Ni120 | 0 to +150°C | 0 to +150°C | 0.1°C | ±(0.02% of rdg +0.1°C) |
| 32 to +302°F | 32 to +302°F | 0.1°F | ± (0.02% of rdg +0.18°F) |
mV | 1 µV | ± (0.01% of rdg +3µV) | ||
mV | 10 µV | ± (0.01% of rdg +3µV) | ||
|
| +21.00 to +53.00 | 10 µV | ± (0.01% of rdg +3µV) |
|
| +53.00 to +100.00 | 10 µV | ± (0.01% of rdg +6µV) |
mV | 0.0 to +1000.0 | 0.0 to +1000.0 | 100 µV | ± (0.01% of rdg +40µV) |
V | 0.000 to +10.000 | 0.000 to +10.000 | 1 mV | ± (0.02% of rdg +0.4mV) |
|
|
|
|
|
mA | 0.000 to +21.000 | 0.000 to +21.000 | 1 µA | ± (0.02% of rdg +0.5 µA) |
Ω (IN) | 0.0 to 400.0 Ω | 0.0 to 400.0 Ω | 10 mΩ | ± (0.02% of rdg +38mΩ) |
Ω (OUT) | 0.0 to 400.0 Ω | 0.0 to 400.0 Ω | 10 mΩ | ± (0.03% of rdg +78mΩ) |
Note:
•Accuracy shown are based on tests at 23°C ±5°C for 360 days
•All Input ranges: additional error ± 1 digit
•Traceability chart to WECC or SIT available on request
9