Omega CL526 manual Scale factor program, RJ 0.0C ext, HI 400.0 Prog, Mode Linear Mode Square

Page 32

RJ : 0.0°C ext

is displayed, it means that an external reference junction compensation has been selected for a temperature of 0.0°C (programmable from -50°C to +100°C).

The reference junction compensation mode can be reprogrammed as indicated in par. 8.5.6. The external reference junction compensation value can be programmed as indicated in par. 8.5.13.2.

9.3.12Scale factor program

The “scale factor” mode is a method to read or to simulate electrical signals values in terms of engineering units.

The example explains the procedure of installing the “scale factor” function for the calibration of a potentiometric recorder with a scale from 0.0 mbar to 400.0 mbar corresponding to the required electrical linear input signal.

Press <SHIFT> + <PROGRAM X> keys to enter the "Scale Factor" set-up procedure. The display will indicate the low end of the scale eg. in mbar.

LO:

0.0 Prog

Press one of the <!> or <"> keys if a decimal point shift is required.

Press the <> or <> keys to adjust to the required value.

Press the <ENTER> key to load in the memory the value and to advance the program one step: the display will indicate the full scale value of the technical unit (eg. mbar).

HI: 400.0 Prog

Press the <▲> or <▼> keys to adjust the full scale value.

Press the <ENTER> key to load in the memory the value: the display will indicate one of the menu pages as follows:

Type : 0-1000 mV

Type : 0-100 mV

Type : 0-10 V

Type : 1-5 V

Type : 0-400

Type : 4-20 mA

Type : 0-20 mA

Select, through <▲> or <▼> keys, the required page and

press the <ENTER> key to memory load the needed parameter. The display will indicate one of the two following pages:

Mode : Linear

Mode : Square

32

Image 32
Contents CL526 OMEGAnetSM On-Line Service Internet e-mail Info@omega.com Introductory Note Table of Contents Digital Interface General Performance Specifications Power supply Battery lifeBattery voltage Dimensions 120x60x230 mm Weights net 1 KgTable of ranges and accuracy 32 to +302F ± 0.02% of rdg +0.18FGeneral Features Case Physical Description + 5 V for logic and analog circuits 5 V for analog circuits Power supplyFunctional Description Input circuit Microprocessor FirmwareDigital display Digital to analog converterBattery charger. Operation from line source Resistance and Rtd measurementsResistance and Rtd simulation Thermocouples input-output circuitUnpacking PRE-OPERATIONAL Check Out 1248.3C TcWiring practice Electrical ConnectionsThermocouple wires Examples of input /output wiring and connectionsLimit of Error of thermocouple wires ClassPower Supply Battery ChargerHow to maximize the battery life Battery Line OpPower on Battery voltage indicationOperating mode set up Operation & ApplicationsOUT mode selection Parameter or sensor selection1032.2C TcK Out 0.0C TcKInternational Temperature Scale Rj modeTecnical unit Decimal point positionIN-OUT data memories Data memory configurationAverage readings Data memory manual recall Data memory automatic scanningSimulation cycle selection Manual step advanceAutomatic simulation cycle Time 0h 0m 50s Start 0.0mVEnd 100.0mV 0mVRj compensation mode check Simulation cycleMode Linear Mode Square Scale factor programRJ 0.0C ext ProgWord mbar Word psi Word hPaExternal Rj compensation Installation parameter procedureFirmware version code Serial number Digital interface data program mode Digital output wiring practiceDigital Interface Baud RateComputer CL526 TTL to RS 232 adapterCommunication protocol from CL526 to a PC Instruct DataPage Ramp Computer request for CL526 settings = OHMInstr ModeDis/Lin STO 2 #D 127 202 Display Chksum = DATA1 + DATA2 + Data 3 + Data 4 .AND.7F Communication programs Example a data transfer from CL526 to PCExample B PC instructions to CL526 Example a3ØØ While LOC 1 = Ø Wend Maintenance Safety recommendationsFaulty operating conditions Error ChecksumProtection fuses StorageIndex WARRANTY/DISCLAIMER Temperature