Omega Multifunction Indicator-Simulator, CL526 OUT mode selection, Parameter or sensor selection

Page 25

9.3.1IN - OUT mode selection

Press the <ON> key to power the instrument.

After diagnostic routine, the calibrator will be forced into the “IN” function with the active parameter previously selected (i.e. with the indication of a measured value of +1032 °C with thermocouple type “K”).

In 1032.2°C TcK:

Open input terminals will cause a fluctuation of the reading up to “Underflow “ or “Overflow“ conditions.

To select the simulation mode press the <IN-OUT> key (the indication will be for example, relative to a simulated value of 0 °C for a thermocouple type “K”).

Out 0.0°C TcK:

The output value can be programmed using the two membrane slidewires (<▲> and <> keys).

Keep the key pressed to cause a continuous variation of the simulated value; the speed of variation will change by moving the pressure to the extremity of the keys.

By touching a point, near the two central zones, the value will increase or decrease by one single digit.

Press simultaneously <> and <> cursor slidewires to set the simulated value to zero.

9.3.2Parameter or sensor selection

To select the electrical parameter or the sensor required by the application, in any measuring or simulation mode, follow the procedure indicated below.

Switch the instrument -ON-

Press the <SELECT> key : the display will show one of the following menu pages:

Tc J K T U L N E

Tc R S B C F G D

Pt100 IEC JIS US

ni100 ni120 Ohm

mV V mA Xscaling

Press <▲> or <▼> cursor slidewire to select the appropriate page.

Select the required parameter or sensor, by moving left or right the flashing cursor with keys <!> or <"> (eg. to activate the thermocouple type T choose the page and cursor position as indicated below).

Tc J K T U L N E

Press the <ENTER> key to memory load the selection; the instrument will return to the previous operative mode with the new selected electrical signal or sensor.

By pressing the <SELECT> key, instead of <ENTER>, the instrument will not acknowledge any variation and return to the previous parameter or sensor.

25

Image 25
Contents CL526 OMEGAnetSM On-Line Service Internet e-mail Info@omega.com Introductory Note Table of Contents Digital Interface General Performance Specifications Battery life Power supplyBattery voltage Dimensions 120x60x230 mm Weights net 1 Kg32 to +302F ± 0.02% of rdg +0.18F Table of ranges and accuracyGeneral Features Case Physical Description Functional Description Power supply+ 5 V for logic and analog circuits 5 V for analog circuits Input circuit Firmware MicroprocessorDigital display Digital to analog converterResistance and Rtd measurements Battery charger. Operation from line sourceThermocouples input-output circuit Resistance and Rtd simulationUnpacking Out 1248.3C Tc PRE-OPERATIONAL CheckElectrical Connections Wiring practiceExamples of input /output wiring and connections Thermocouple wires Class Limit of Error of thermocouple wiresBattery Charger Power SupplyHow to maximize the battery life Battery Line OpBattery voltage indication Power onOperating mode set up Operation & ApplicationsParameter or sensor selection OUT mode selection1032.2C TcK Out 0.0C TcKRj mode International Temperature ScaleTecnical unit Decimal point positionAverage readings Data memory configurationIN-OUT data memories Data memory automatic scanning Data memory manual recallAutomatic simulation cycle Manual step advanceSimulation cycle selection Start 0.0mV Time 0h 0m 50sEnd 100.0mV 0mVSimulation cycle Rj compensation mode checkScale factor program Mode Linear Mode SquareRJ 0.0C ext ProgWord psi Word hPa Word mbarFirmware version code Serial number Installation parameter procedureExternal Rj compensation Digital output wiring practice Digital interface data program modeDigital Interface Baud RateCommunication protocol from CL526 to a PC TTL to RS 232 adapterComputer CL526 Data InstructPage Ramp = OHM Computer request for CL526 settingsMode InstrDis/Lin STO 2 #D 127 202 Display Chksum = DATA1 + DATA2 + Data 3 + Data 4 .AND.7F Example a data transfer from CL526 to PC Communication programsExample B PC instructions to CL526 Example a3ØØ While LOC 1 = Ø Wend Safety recommendations MaintenanceFaulty operating conditions Error ChecksumStorage Protection fusesIndex WARRANTY/DISCLAIMER Temperature