To select the required group follow the procedure indicated below:
•Press the <SELECT> key to obtain one of the menu pages
•Press the <0> key to obtain the following indication:
Sel STO Group: A
•Press the <▲> or <▼> key to select the required group number
•Press the <0> key to confirm the selection and to return to the previously selected operative mode.
9.3.9.2Data memory manual recall
To recall data memory values, select first the appropriate or required group number and press then the <0>, <1>, or <2> key.
9.3.9.3Data memory automatic scanning
The 60 stored items of data can be assigned, besides the group number, to a
The programmed sequence (a number of calibration points, or a number of data items to be supervised), includes a linear sequence of memory starting from a "start" point and ending with a "end" point".
Memory for seven different
The numerical value and the parameter indicated below are an operative example eg. to memory load 5 calibration points of a scale of a potentiometric recorder (temperature with a thermocouple type K).
1 =
2 = 0°C
3 = +100°C
4 = +200°C
5 = +500°C
As described in par. 8.6.9.1 memory store the calibration data as follows:
point 1 = memory 0 - group A point 2 = memory 1 - group A point 3 = memory 2 - group A point 4 = memory 0 - group B point 5 = memory 1 - group B
The simulation program
•Press the <SELECT> key to obtain one of the menu pages
•Press the <1> key to obtain the following indication:
SEL Program # 1
•Press the <▲> and <▼> keys to enter the program number required (1 to 7)
•Press the <1> key to confirm the selection of the program number and to return to the indication of one of the menu pages
•Press the <2> key to obtain the following indication:
Prog #1 from A1
•Press the <▲> and <▼> keys to select the "start" group/memory (eg. A1)
•Press the <2> key to confirm the "start" point obtaining the following indication:
Prog #1 | to A2 |
•Press <s> and <t> keys to select the "end" group/memory (eg. A2)
28