7 ELECTRICAL CONNECTIONS

Appropriate extension wires should be used between the thermocouple (or instrument under calibration) and the CL526 unless the thermocouple leads permit direct connection.

Make sure that both thermocouple and compensating cable are connected with the correct polarity.

If in doubt, the polarity of the compensating leads can be checked by connecting a length of lead to the indicator, shorting the free ends of the wires together and noting that the indicator reading increases when the wires connection is heated.

Color codes of compensating cables change in different countries. Check the appropriate table. For Rtd connection use a cable of adequate gauge to lower the overall input resistance.

The use of a cable with a good resistance balance between conductors is also necessary.

Table A

Colour code & polarity for extension wires

Thermocouple

EChromel Constantan

Iron

J

Constantan

Chromel

K

Alumel

Pt 13% Rh

R

Platinum

Pt 10% Rh

S

Platinum

Copper

T

Constantan

Pt 6% Rh

B

Pt 30% Rh

Nicrosil

N

Nisil

 

 

 

 

Wires

 

 

 

 

 

 

 

 

( + )

 

 

Chromel

( -

 

)

 

Constantan

( + )

 

 

Iron

(

-

 

) Constantan

 

 

 

 

( + )

 

 

Chromel

(

-

 

) Alumel

( + )

 

 

Copper

( -

 

)

 

Alloy 11

 

 

 

 

( + )

 

 

Copper

( -

 

)

 

Alloy 11

( + )

 

 

Copper

( -

 

)

 

Constantan

( + )

 

 

Copper

( -

 

)

 

Copper

 

 

 

 

( + )

 

 

Nicrosil

( -

 

)

 

Nisil

Colour code

Purple Red

White

Red

Yellow Red

Black

Red

Black

Red

Blue

Red

Orange Red

7.1Wiring practice

Although the CL526 portable calibrator is designed to be insensitive to transients or noise, the following recommendations should be followed to reduce ac pick up in the signal leads and to ensure good performance.

The input leads should not be run near ac line wiring, transformers and heating elements.

Input/output leads should, if possible, be twisted and shielded with the shield grounded at the end of the cable. When shielded cables are used the shield must be connected to the positive terminal.

Above figure shows some examples of input/output wiring and connections.

20

Image 20
Contents CL526 OMEGAnetSM On-Line Service Internet e-mail Info@omega.com Introductory Note Table of Contents Digital Interface General Performance Specifications Power supply Battery lifeBattery voltage Dimensions 120x60x230 mm Weights net 1 KgTable of ranges and accuracy 32 to +302F ± 0.02% of rdg +0.18FGeneral Features Case Physical Description + 5 V for logic and analog circuits 5 V for analog circuits Power supplyFunctional Description Input circuit Microprocessor FirmwareDigital display Digital to analog converterBattery charger. Operation from line source Resistance and Rtd measurementsResistance and Rtd simulation Thermocouples input-output circuitUnpacking PRE-OPERATIONAL Check Out 1248.3C TcWiring practice Electrical ConnectionsThermocouple wires Examples of input /output wiring and connectionsLimit of Error of thermocouple wires ClassPower Supply Battery ChargerHow to maximize the battery life Battery Line OpPower on Battery voltage indicationOperating mode set up Operation & ApplicationsOUT mode selection Parameter or sensor selection1032.2C TcK Out 0.0C TcKInternational Temperature Scale Rj modeTecnical unit Decimal point positionIN-OUT data memories Data memory configurationAverage readings Data memory manual recall Data memory automatic scanningSimulation cycle selection Manual step advanceAutomatic simulation cycle Time 0h 0m 50s Start 0.0mVEnd 100.0mV 0mVRj compensation mode check Simulation cycleMode Linear Mode Square Scale factor programRJ 0.0C ext ProgWord mbar Word psi Word hPaExternal Rj compensation Installation parameter procedureFirmware version code Serial number Digital interface data program mode Digital output wiring practiceDigital Interface Baud RateComputer CL526 TTL to RS 232 adapterCommunication protocol from CL526 to a PC Instruct DataPage Ramp Computer request for CL526 settings = OHMInstr ModeDis/Lin STO 2 #D 127 202 Display Chksum = DATA1 + DATA2 + Data 3 + Data 4 .AND.7F Communication programs Example a data transfer from CL526 to PCExample B PC instructions to CL526 Example a3ØØ While LOC 1 = Ø Wend Maintenance Safety recommendationsFaulty operating conditions Error ChecksumProtection fuses StorageIndex WARRANTY/DISCLAIMER Temperature