Omega CL526 Operation & Applications, Power on, Battery voltage indication, Operating mode set up

Page 24
Press the <A> key and keeping the pressure on the key, press then the <B> key. Press in sequence, first the <A> key and then the <B> key.

9 OPERATION & APPLICATIONS

9.1Power ON

ATTENTION: ALL VALUES IN THE FOLLOWING FIGURES ARE ONLY LISTED AS AN EXAMPLE.

During set-up and load memory remember that the instructions of the manual related to key operation have the following meaning:

<A> + <B> <A> , <B>

To power the instrument press the <ON> key; the indication :

... CL526...

will appear for a few seconds.

The instrument runs an autodiagnostic routine for the self-checking of critical circuits and components.

A positive check will be shown with the indication

Test OK Ver 7.001

for about one second.

The number on the right side of the display indicates the version of the memory installed on the instrument.

The instrument is ready for measurement (IN mode) with the previously selected operating mode, as indicated below:

In 1280.6°C TcK:

Any faulty conditions that may be indicated are described in par. 8.9.

9.2Battery voltage indication

To recall the battery voltage on the display press the <SHIFT> + <BATTERY> keys. The indication will be as follows:

BAT 5.2 V IIII

The horizontal bar indicates the level of charge of the battery (each bar is equivalent to 25% of the full capacity)

The “low” limit of the battery voltage, for the correct operation of the instrument, is +4.6V.

Press any key to reset the operative mode.

During normal operating modes (measure or simulation), “low battery” condition will be shown as follows:

Out 1248.3°C TcK:

The battery symbol indicates that the battery has enough energy for about 30 minutes operation. A black symbol means that battery charge is below minimum: batteries must be recharged.

9.3Operating mode set up

To select the required operating mode follow the procedures indicated below.

24

Image 24
Contents CL526 OMEGAnetSM On-Line Service Internet e-mail Info@omega.com Introductory Note Table of Contents Digital Interface General Performance Specifications Power supply Battery lifeBattery voltage Dimensions 120x60x230 mm Weights net 1 KgTable of ranges and accuracy 32 to +302F ± 0.02% of rdg +0.18FGeneral Features Case Physical Description Power supply Functional Description+ 5 V for logic and analog circuits 5 V for analog circuits Input circuit Microprocessor FirmwareDigital display Digital to analog converterBattery charger. Operation from line source Resistance and Rtd measurementsResistance and Rtd simulation Thermocouples input-output circuitUnpacking PRE-OPERATIONAL Check Out 1248.3C TcWiring practice Electrical ConnectionsThermocouple wires Examples of input /output wiring and connectionsLimit of Error of thermocouple wires ClassPower Supply Battery ChargerHow to maximize the battery life Battery Line OpPower on Battery voltage indicationOperating mode set up Operation & ApplicationsOUT mode selection Parameter or sensor selection1032.2C TcK Out 0.0C TcKInternational Temperature Scale Rj modeTecnical unit Decimal point positionData memory configuration Average readingsIN-OUT data memories Data memory manual recall Data memory automatic scanningManual step advance Automatic simulation cycleSimulation cycle selection Time 0h 0m 50s Start 0.0mVEnd 100.0mV 0mVRj compensation mode check Simulation cycleMode Linear Mode Square Scale factor programRJ 0.0C ext ProgWord mbar Word psi Word hPaInstallation parameter procedure Firmware version code Serial numberExternal Rj compensation Digital interface data program mode Digital output wiring practiceDigital Interface Baud RateTTL to RS 232 adapter Communication protocol from CL526 to a PCComputer CL526 Instruct DataPage Ramp Computer request for CL526 settings = OHMInstr ModeDis/Lin STO 2 #D 127 202 Display Chksum = DATA1 + DATA2 + Data 3 + Data 4 .AND.7F Communication programs Example a data transfer from CL526 to PCExample B PC instructions to CL526 Example a3ØØ While LOC 1 = Ø Wend Maintenance Safety recommendationsFaulty operating conditions Error ChecksumProtection fuses StorageIndex WARRANTY/DISCLAIMER Temperature