Omega Multifunction Indicator-Simulator, CL526 manual Protection fuses, Storage

Page 47

ERROR

6

When in mV or Tc “OUT” function the current flow must not

exceed 0.5 mA.

ERROR

7

Indicates a possible error during scale factor computation.

 

The above indicated faulty conditions can be announced both during the autodiagnostic routine or in measure or simulation modes.

If the faulty condition is critical for the type of application, it is recommended to re-run the pertinent set up procedure.

11.3Protection fuses

The CL526 unit, is protected against overvoltage and overcurrent inputs, by thermal fuses. When they activate, you should disconnect the electrical connections from the unit and switch the instrument off for about 2 minutes. The thermal fuses will be automatically restored and you can resume your work.

11.4Storage

If the instrument is left unused for a long time, it is recommended to remove the batteries.

Store the instrument in the original package, at a temperature from -30°C to +60°C, with R.H. less than 90%.

If the instrument has been unused for a month check the battery voltage, and charge the Ni-MH batteries for at least 12 hours.

47

Image 47
Contents CL526 OMEGAnetSM On-Line Service Internet e-mail Info@omega.com Introductory Note Table of Contents Digital Interface General Performance Specifications Dimensions 120x60x230 mm Weights net 1 Kg Power supplyBattery life Battery voltage32 to +302F ± 0.02% of rdg +0.18F Table of ranges and accuracyGeneral Features Case Physical Description + 5 V for logic and analog circuits 5 V for analog circuits Power supplyFunctional Description Input circuit Digital to analog converter MicroprocessorFirmware Digital displayResistance and Rtd measurements Battery charger. Operation from line sourceThermocouples input-output circuit Resistance and Rtd simulationUnpacking Out 1248.3C Tc PRE-OPERATIONAL CheckElectrical Connections Wiring practiceExamples of input /output wiring and connections Thermocouple wiresClass Limit of Error of thermocouple wiresBattery Line Op Power SupplyBattery Charger How to maximize the battery lifeOperation & Applications Power onBattery voltage indication Operating mode set upOut 0.0C TcK OUT mode selectionParameter or sensor selection 1032.2C TcKDecimal point position International Temperature ScaleRj mode Tecnical unitIN-OUT data memories Data memory configurationAverage readings Data memory automatic scanning Data memory manual recallSimulation cycle selection Manual step advanceAutomatic simulation cycle 0mV Time 0h 0m 50sStart 0.0mV End 100.0mVSimulation cycle Rj compensation mode checkProg Mode Linear Mode SquareScale factor program RJ 0.0C extWord psi Word hPa Word mbarExternal Rj compensation Installation parameter procedureFirmware version code Serial number Baud Rate Digital interface data program modeDigital output wiring practice Digital InterfaceComputer CL526 TTL to RS 232 adapterCommunication protocol from CL526 to a PC Data InstructPage Ramp = OHM Computer request for CL526 settingsMode InstrDis/Lin STO 2 #D 127 202 Display Chksum = DATA1 + DATA2 + Data 3 + Data 4 .AND.7F Example a Communication programs Example a data transfer from CL526 to PC Example B PC instructions to CL5263ØØ While LOC 1 = Ø Wend Error Checksum MaintenanceSafety recommendations Faulty operating conditionsStorage Protection fusesIndex WARRANTY/DISCLAIMER Temperature