Agilent Technologies 8753ET, ES quick start Configure the Analyzer

Page 19

Installing Your Analyzer

STEP 4. Configure the Analyzer

5.If you will be using the serial port, adjust the analyzer's baud rate until it is equal to the baud rate set on the peripheral by pressing PLOTTER BAUD RATE or

PRINTER BAUD RATE and the and front panel keys.

NOTE

The plotter menu is shown as an example. It will only appear if you select

 

 

.

 

PLOTTER PORT

 

 

 

You can set the analyzer to the following baud rates:

• 1200

• 2400

• 4800

• 9600

• 19200

6.Also, if you will be using the serial port, you must toggle the transmission control

XMIT CNTRL (handshaking protocol) until your choice of Xon/Xoff or DTR/DSR appears (equal to the transmission control set on the peripheral). The printer menu is

shown as an example. It will only appear if you select PRINTER PORT .

NOTE

Transmission control for plotters is set programmatically.

Xon/Xoff sets transmission on/transmission off (software handshake).

DTR/DSR sets data terminal ready/data set ready (hardware handshake).

Chapter 1

1-13

Image 19
Contents Part Number Print Date February Installation and Quick Start GuidePage Certification General Safety Considerations How to Use This Guide Documentation Map Contents Installing Your Analyzer Introduction Verify the Shipment Item Number Part Number Description Analyzer Front Panel Familiarize Yourself with the Analyzer Front and Rear PanelsTest Sequence Analyzer Rear PanelMeet Electrical and Environmental Requirements Meet Electrical and Environmental Requirements Configure the Analyzer To Configure the Standard Analyzer To Configure the Analyzer with Printers or Plotters Plotter Port Configure the Analyzer Thinkjet Deskjet SET Clock To Configure the Analyzer for Bench Top or Rack Mount Use To Attach Front Handles to the Analyzer Standard Configure the Analyzer Configure the Analyzer Verify the Analyzer Operation To View the Installed Options To Initiate the Analyzer Self-Test To Run the Operators Check Service Menu TestsTrans FWD S21B/R or Transmissn To Test the Transmission ModeTo Test the Reflection Mode Equipment Eeprom Backup Disk ProcedureBack Up the Eeprom Disk DescriptionBack Up the Eeprom Disk Quick Start Learning How to Make Measurements Quick Start Learning How to Make Measurements Analyzer Front Panel Analyzer Front PanelAnalyzer Front Panel Output measurement results Measurement ProcedureMake a measurement calibration Measure the deviceLearning to Make Transmission Measurements Choose the following measurement settingsConnections for a Thru Calibration Standard Perform a measurement calibrationExample Measurement of Insertion Loss Press Return Define DISK-SAVE Example Measurement of 3 dB Bandwidth Measuring Other Transmission CharacteristicsLearning to Make Transmission Measurements Example Measurement of Out-of-Band Rejection Measuring Out-of-Band RejectionExample Measurement of Passband Flatness or Ripple Measuring Passband Flatness or Ripple= Γ Learning to Make Reflection Measurements= 1 + ρ / 1 − ρ Connections for Reflection Measurements Make a measurement calibration 10 Example Measurement of Return Loss Measuring Return Loss11 Example Measurement Title Measuring Reflection Coefficient Measuring Other Reflection Characteristics13 Example Standing-Wave-Ratio Measurement Trace Measuring Standing Wave Ratio SWRMeasuring S11 and S22 or Reflection in a Polar Format Measuring Impedance 15 Example Impedance Measurement Trace 16 Example Admittance Measurement Trace Power-Up Problems If You Encounter a ProblemIf the analyzer display does not light If there is no RF signal at the front-panel port Data Entry ProblemsNo RF Output If You Encounter a Problem European Field Operations Agilent Technologies Sales and Service OfficesIndex Index
Related manuals
Manual 28 pages 56.53 Kb Manual 61 pages 32.91 Kb