This example procedure shows you how to measure the transmission response of a 175 MHz bandpass filter. The measurement parameters listed are unique to this particular test device.

For further measurement examples, refer to the “Making Measurements” chapter in the User's Guide.

Step 1. Choose the measurement parameters with your test device connected

1.Press the Preset key to return the analyzer to a known state.

2.Connect your test device to the analyzer as shown in Figure 2-2.Use adapters where appropriate.

CAUTION Damage may result to the device under test if it is sensitive to the analyzer's default output power level. To avoid damaging a sensitive DUT, be sure to set the analyzer's output power to an appropriate level before connecting the DUT to the analyzer.

Figure 2-2 Device Connections for a Transmission Measurement

3.Choose the following measurement settings:

Meas Trans: FWD S21 (B/R) or TRANSMISSN

Center200 M/μ

Span375 M/μ

Power5x1

Scale Ref AUTO SCALE

4.Look at the device response to determine if these are the parameters that you want for your device measurement. For example, if the trace is noisy you may want to increase the test port output power (which increases the analyzer input power), reduce the IF bandwidth, or add averaging. Or, to better see an area of interest, you may want to change the test frequencies.

2-6

Chapter 2

Image 39
Contents Part Number Print Date February Installation and Quick Start GuidePage Certification General Safety Considerations How to Use This Guide Documentation Map Contents Installing Your Analyzer Introduction Verify the Shipment Item Number Part Number Description Analyzer Front Panel Familiarize Yourself with the Analyzer Front and Rear PanelsTest Sequence Analyzer Rear PanelMeet Electrical and Environmental Requirements Meet Electrical and Environmental Requirements Configure the Analyzer To Configure the Standard Analyzer To Configure the Analyzer with Printers or Plotters Plotter Port Configure the Analyzer Thinkjet Deskjet SET Clock To Configure the Analyzer for Bench Top or Rack Mount Use To Attach Front Handles to the Analyzer Standard Configure the Analyzer Configure the Analyzer Verify the Analyzer Operation To View the Installed Options To Initiate the Analyzer Self-Test To Run the Operators Check Service Menu TestsTrans FWD S21B/R or Transmissn To Test the Transmission ModeTo Test the Reflection Mode Equipment Eeprom Backup Disk ProcedureBack Up the Eeprom Disk DescriptionBack Up the Eeprom Disk Quick Start Learning How to Make Measurements Quick Start Learning How to Make Measurements Analyzer Front Panel Analyzer Front PanelAnalyzer Front Panel Output measurement results Measurement ProcedureMake a measurement calibration Measure the deviceLearning to Make Transmission Measurements Choose the following measurement settingsConnections for a Thru Calibration Standard Perform a measurement calibrationExample Measurement of Insertion Loss Press Return Define DISK-SAVE Example Measurement of 3 dB Bandwidth Measuring Other Transmission CharacteristicsLearning to Make Transmission Measurements Example Measurement of Out-of-Band Rejection Measuring Out-of-Band RejectionExample Measurement of Passband Flatness or Ripple Measuring Passband Flatness or RippleLearning to Make Reflection Measurements = Γ= 1 + ρ / 1 − ρ Connections for Reflection Measurements Make a measurement calibration 10 Example Measurement of Return Loss Measuring Return Loss11 Example Measurement Title Measuring Reflection Coefficient Measuring Other Reflection Characteristics13 Example Standing-Wave-Ratio Measurement Trace Measuring Standing Wave Ratio SWRMeasuring S11 and S22 or Reflection in a Polar Format Measuring Impedance 15 Example Impedance Measurement Trace 16 Example Admittance Measurement Trace If You Encounter a Problem Power-Up ProblemsIf the analyzer display does not light Data Entry Problems If there is no RF signal at the front-panel portNo RF Output If You Encounter a Problem European Field Operations Agilent Technologies Sales and Service OfficesIndex Index