Agilent Technologies 8753ET, ES quick start Meet Electrical and Environmental Requirements

Page 13

Installing Your Analyzer

STEP 3. Meet Electrical and Environmental Requirements

STEP 3. Meet Electrical and Environmental Requirements

1. Set the line-voltage selector to the position

 

2. Ensure the available AC power source

 

that corresponds to the AC power source.

 

 

meets the following requirements:

 

 

 

 

 

 

 

 

90–132 VAC

 

 

 

47–66 Hz / 400 Hz (single phase)

 

 

 

 

- or -

 

 

 

 

198–265 VAC

 

 

 

47–66 Hz (single phase)

 

 

 

The analyzer power consumption is 350 VA

 

 

 

maximum.

 

 

 

 

 

 

3. Ensure the operating environment meets

 

4. Verify that the power cable is not damaged,

 

the following requirements:

 

 

and that the power-source outlet provides a

 

 

 

 

protective earth contact.

 

 

 

 

 

 

 

• 0 to 55 °C

 

 

 

 

 

• < 95% relative humidity at 40 °C

 

 

 

 

 

(non-condensing)

 

 

 

 

 

• < 15,000 feet (≈ 4,500 meters) altitude

 

 

 

 

 

Some analyzer performance parameters are

 

 

 

 

 

specified for 25 °C ±5 °C. Refer to the

 

 

 

 

 

Reference Guide for information on the

 

 

 

 

 

environmental compatibility of warranted

 

 

 

 

 

performance.

 

 

 

 

 

 

 

 

 

 

 

 

WARNING

Any interruption of the

 

 

 

 

 

protective (grounding)

 

 

 

 

 

conductor or disconnection

 

 

 

 

 

of the protective earth

 

 

 

 

 

terminal, can result in

 

 

 

 

 

personal injury, or may

 

 

 

 

 

damage the analyzer.

 

 

 

 

 

 

 

 

 

 

 

 

 

Chapter 1

1-7

Image 13
Contents Part Number Print Date February Installation and Quick Start GuidePage Certification General Safety Considerations How to Use This Guide Documentation Map Contents Installing Your Analyzer Introduction Verify the Shipment Item Number Part Number Description Analyzer Front Panel Familiarize Yourself with the Analyzer Front and Rear PanelsTest Sequence Analyzer Rear PanelMeet Electrical and Environmental Requirements Meet Electrical and Environmental Requirements Configure the Analyzer To Configure the Standard Analyzer To Configure the Analyzer with Printers or Plotters Plotter Port Configure the Analyzer Thinkjet Deskjet SET Clock To Configure the Analyzer for Bench Top or Rack Mount Use To Attach Front Handles to the Analyzer Standard Configure the Analyzer Configure the Analyzer Verify the Analyzer Operation To View the Installed Options To Initiate the Analyzer Self-Test To Run the Operators Check Service Menu TestsTrans FWD S21B/R or Transmissn To Test the Transmission ModeTo Test the Reflection Mode Back Up the Eeprom Disk Eeprom Backup Disk ProcedureDescription EquipmentBack Up the Eeprom Disk Quick Start Learning How to Make Measurements Quick Start Learning How to Make Measurements Analyzer Front Panel Analyzer Front PanelAnalyzer Front Panel Make a measurement calibration Measurement ProcedureMeasure the device Output measurement resultsLearning to Make Transmission Measurements Choose the following measurement settingsConnections for a Thru Calibration Standard Perform a measurement calibrationExample Measurement of Insertion Loss Press Return Define DISK-SAVE Example Measurement of 3 dB Bandwidth Measuring Other Transmission CharacteristicsLearning to Make Transmission Measurements Example Measurement of Out-of-Band Rejection Measuring Out-of-Band RejectionExample Measurement of Passband Flatness or Ripple Measuring Passband Flatness or Ripple= Γ Learning to Make Reflection Measurements= 1 + ρ / 1 − ρ Connections for Reflection Measurements Make a measurement calibration 10 Example Measurement of Return Loss Measuring Return Loss11 Example Measurement Title Measuring Reflection Coefficient Measuring Other Reflection Characteristics13 Example Standing-Wave-Ratio Measurement Trace Measuring Standing Wave Ratio SWRMeasuring S11 and S22 or Reflection in a Polar Format Measuring Impedance 15 Example Impedance Measurement Trace 16 Example Admittance Measurement Trace Power-Up Problems If You Encounter a ProblemIf the analyzer display does not light If there is no RF signal at the front-panel port Data Entry ProblemsNo RF Output If You Encounter a Problem European Field Operations Agilent Technologies Sales and Service OfficesIndex Index