Agilent Technologies 8753ET, ES quick start Back Up the Eeprom Disk

Page 33

Installing Your Analyzer

STEP 6. Back Up the EEPROM Disk

5. Press FILE UTILITIES RENAME FILE ERASE TITLE . Use the front panel knob

and the SELECT LETTER softkey to rename the file “FILE00” to “N12345” where 12345 represents the last 5 digits of the instrument's serial number. (The first character in the file name must be a letter.) When finished, press DONE .

6.Label the disk with the serial number of the instrument, the date, and the words “EEPROM Backup Disk.”

NOTE

Whenever the analyzer is returned to Agilent Technologies for servicing

 

and/or calibration, the EEPROM backup disk should be returned with the

 

analyzer. This will significantly reduce the instrument repair time.

 

 

7. The EEPROM backup disk procedure is now complete.

Chapter 1

1-27

Image 33
Contents Part Number Print Date February Installation and Quick Start GuidePage Certification General Safety Considerations How to Use This Guide Documentation Map Contents Installing Your Analyzer Introduction Verify the Shipment Item Number Part Number Description Analyzer Front Panel Familiarize Yourself with the Analyzer Front and Rear PanelsTest Sequence Analyzer Rear PanelMeet Electrical and Environmental Requirements Meet Electrical and Environmental Requirements Configure the Analyzer To Configure the Standard Analyzer To Configure the Analyzer with Printers or Plotters Plotter Port Configure the Analyzer Thinkjet Deskjet SET Clock To Configure the Analyzer for Bench Top or Rack Mount Use To Attach Front Handles to the Analyzer Standard Configure the Analyzer Configure the Analyzer Verify the Analyzer Operation To View the Installed Options To Initiate the Analyzer Self-Test To Run the Operators Check Service Menu TestsTrans FWD S21B/R or Transmissn To Test the Transmission ModeTo Test the Reflection Mode Back Up the Eeprom Disk Eeprom Backup Disk ProcedureDescription EquipmentBack Up the Eeprom Disk Quick Start Learning How to Make Measurements Quick Start Learning How to Make Measurements Analyzer Front Panel Analyzer Front PanelAnalyzer Front Panel Make a measurement calibration Measurement ProcedureMeasure the device Output measurement resultsLearning to Make Transmission Measurements Choose the following measurement settingsConnections for a Thru Calibration Standard Perform a measurement calibrationExample Measurement of Insertion Loss Press Return Define DISK-SAVE Example Measurement of 3 dB Bandwidth Measuring Other Transmission CharacteristicsLearning to Make Transmission Measurements Example Measurement of Out-of-Band Rejection Measuring Out-of-Band RejectionExample Measurement of Passband Flatness or Ripple Measuring Passband Flatness or RippleLearning to Make Reflection Measurements = Γ= 1 + ρ / 1 − ρ Connections for Reflection Measurements Make a measurement calibration 10 Example Measurement of Return Loss Measuring Return Loss11 Example Measurement Title Measuring Reflection Coefficient Measuring Other Reflection Characteristics13 Example Standing-Wave-Ratio Measurement Trace Measuring Standing Wave Ratio SWRMeasuring S11 and S22 or Reflection in a Polar Format Measuring Impedance 15 Example Impedance Measurement Trace 16 Example Admittance Measurement Trace If You Encounter a Problem Power-Up ProblemsIf the analyzer display does not light Data Entry Problems If there is no RF signal at the front-panel portNo RF Output If You Encounter a Problem European Field Operations Agilent Technologies Sales and Service OfficesIndex Index
Related manuals
Manual 28 pages 56.53 Kb Manual 61 pages 32.91 Kb