Agilent Technologies ES Measurement Procedure, Make a measurement calibration, Measure the device

Page 38

Quick Start: Learning How to Make Measurements

Measurement Procedure

Measurement Procedure

This is a general measurement procedure that is used throughout the guide to illustrate the use of the analyzer.

Step 1. Choose measurement parameters with your test device connected

Press the Preset key to return the analyzer to a known state.

Connect your device under test (DUT) to the analyzer.

CAUTION Damage may result to the DUT if it is sensitive to the analyzer's default output power level. To avoid damaging a sensitive DUT, be sure to set the analyzer's output power to an appropriate level before connecting the DUT to the analyzer.

Choose the settings that are appropriate for the intended measurement.

measurement type (S11 or reflection, for example)

frequencies

number of points

power

measurement trace format

Make adjustments to the parameters while you are viewing the device response.

Step 2. Make a measurement calibration

Press the Cal key to begin to perform a measurement calibration using a known set of standards (a calibration kit). Error-correction establishes a magnitude and phase reference for the test setup and reduces systematic measurement errors.

Step 3. Measure the device

Reconnect the device under test.

Use the markers to identify various device response values if desired.

Step 4. Output measurement results

Store the measurement file to a disk.

Generate a hardcopy with a printer or plotter.

Chapter 2

2-5

Image 38
Contents Installation and Quick Start Guide Part Number Print Date FebruaryPage Certification General Safety Considerations How to Use This Guide Documentation Map Contents Installing Your Analyzer Introduction Verify the Shipment Item Number Part Number Description Familiarize Yourself with the Analyzer Front and Rear Panels Analyzer Front PanelAnalyzer Rear Panel Test SequenceMeet Electrical and Environmental Requirements Meet Electrical and Environmental Requirements Configure the Analyzer To Configure the Standard Analyzer To Configure the Analyzer with Printers or Plotters Plotter Port Configure the Analyzer Thinkjet Deskjet SET Clock To Configure the Analyzer for Bench Top or Rack Mount Use To Attach Front Handles to the Analyzer Standard Configure the Analyzer Configure the Analyzer Verify the Analyzer Operation To View the Installed Options To Initiate the Analyzer Self-Test Service Menu Tests To Run the Operators CheckTo Test the Transmission Mode Trans FWD S21B/R or TransmissnTo Test the Reflection Mode Description Eeprom Backup Disk ProcedureBack Up the Eeprom Disk EquipmentBack Up the Eeprom Disk Quick Start Learning How to Make Measurements Quick Start Learning How to Make Measurements Analyzer Front Panel Analyzer Front PanelAnalyzer Front Panel Measure the device Measurement ProcedureMake a measurement calibration Output measurement resultsChoose the following measurement settings Learning to Make Transmission MeasurementsPerform a measurement calibration Connections for a Thru Calibration StandardExample Measurement of Insertion Loss Press Return Define DISK-SAVE Measuring Other Transmission Characteristics Example Measurement of 3 dB BandwidthLearning to Make Transmission Measurements Measuring Out-of-Band Rejection Example Measurement of Out-of-Band RejectionMeasuring Passband Flatness or Ripple Example Measurement of Passband Flatness or Ripple= 1 + ρ / 1 − ρ Learning to Make Reflection Measurements= Γ Connections for Reflection Measurements Make a measurement calibration Measuring Return Loss 10 Example Measurement of Return Loss11 Example Measurement Title Measuring Other Reflection Characteristics Measuring Reflection CoefficientMeasuring Standing Wave Ratio SWR 13 Example Standing-Wave-Ratio Measurement TraceMeasuring S11 and S22 or Reflection in a Polar Format Measuring Impedance 15 Example Impedance Measurement Trace 16 Example Admittance Measurement Trace If the analyzer display does not light If You Encounter a ProblemPower-Up Problems No RF Output Data Entry ProblemsIf there is no RF signal at the front-panel port If You Encounter a Problem Agilent Technologies Sales and Service Offices European Field OperationsIndex Index