Agilent Technologies ES, 8753ET To Test the Transmission Mode, Trans FWD S21B/R or Transmissn

Page 30

Installing Your Analyzer

STEP 5. Verify the Analyzer Operation

To Test the Transmission Mode

1. Connect the equipment as shown and press

2. Check the forward transmission mode for

Preset .

channel 2 by pressing Chan 2 Meas

Trans: FWD S21(B/R) or TRANSMISSN .

 

NOTE

The test port return cable

 

 

 

 

 

should have low-loss

 

 

 

 

 

characteristics to avoid a

 

 

 

 

 

degradation in frequency

 

 

 

 

 

response at higher frequencies.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3. Look at the measurement trace displayed

4. ES models only: To check the reverse

 

on the analyzer. It should be similar to the

transmission mode for channel 2, press

 

trace below.

 

Meas

 

. The

 

 

Trans: REV S12 (A/R)

 

 

 

measurement trace should be similar to the

 

 

 

trace below.

1-24

Chapter 1

Image 30
Contents Installation and Quick Start Guide Part Number Print Date FebruaryPage Certification General Safety Considerations How to Use This Guide Documentation Map Contents Installing Your Analyzer Introduction Verify the Shipment Item Number Part Number Description Familiarize Yourself with the Analyzer Front and Rear Panels Analyzer Front PanelAnalyzer Rear Panel Test SequenceMeet Electrical and Environmental Requirements Meet Electrical and Environmental Requirements Configure the Analyzer To Configure the Standard Analyzer To Configure the Analyzer with Printers or Plotters Plotter Port Configure the Analyzer Thinkjet Deskjet SET Clock To Configure the Analyzer for Bench Top or Rack Mount Use To Attach Front Handles to the Analyzer Standard Configure the Analyzer Configure the Analyzer Verify the Analyzer Operation To View the Installed Options To Initiate the Analyzer Self-Test Service Menu Tests To Run the Operators CheckTo Test the Transmission Mode Trans FWD S21B/R or TransmissnTo Test the Reflection Mode Description Eeprom Backup Disk ProcedureBack Up the Eeprom Disk EquipmentBack Up the Eeprom Disk Quick Start Learning How to Make Measurements Quick Start Learning How to Make Measurements Analyzer Front Panel Analyzer Front PanelAnalyzer Front Panel Measure the device Measurement ProcedureMake a measurement calibration Output measurement resultsChoose the following measurement settings Learning to Make Transmission MeasurementsPerform a measurement calibration Connections for a Thru Calibration StandardExample Measurement of Insertion Loss Press Return Define DISK-SAVE Measuring Other Transmission Characteristics Example Measurement of 3 dB BandwidthLearning to Make Transmission Measurements Measuring Out-of-Band Rejection Example Measurement of Out-of-Band RejectionMeasuring Passband Flatness or Ripple Example Measurement of Passband Flatness or RippleLearning to Make Reflection Measurements = Γ= 1 + ρ / 1 − ρ Connections for Reflection Measurements Make a measurement calibration Measuring Return Loss 10 Example Measurement of Return Loss11 Example Measurement Title Measuring Other Reflection Characteristics Measuring Reflection CoefficientMeasuring Standing Wave Ratio SWR 13 Example Standing-Wave-Ratio Measurement TraceMeasuring S11 and S22 or Reflection in a Polar Format Measuring Impedance 15 Example Impedance Measurement Trace 16 Example Admittance Measurement Trace If You Encounter a Problem Power-Up ProblemsIf the analyzer display does not light Data Entry Problems If there is no RF signal at the front-panel portNo RF Output If You Encounter a Problem Agilent Technologies Sales and Service Offices European Field OperationsIndex Index
Related manuals
Manual 28 pages 56.53 Kb Manual 61 pages 32.91 Kb