Fluke 80 Series V manual Digital Multimeter

Page 11

Digital Multimeter

Precautions and Safety Information

When making electrical connections, connect the common test lead before connecting the live test lead; when disconnecting, disconnect the live test lead before disconnecting the common test lead.

Do not use the Meter if it operates abnormally. Protection may be impaired. When in doubt, have the Meter serviced.

Do not operate the Meter around explosive gas, vapor, or dust.

Use only a single 9 V battery, properly installed in the Meter case, to power the Meter.

When servicing the Meter, use only specified replacement parts.

When using probes, keep fingers behind the finger guards on the probes.

Do not use the Low Pass Filter option to verify the presence of hazardous voltages. Voltages greater than what is indicated may be present. Make a voltage measurement without the filter to detect the possible presence of hazardous voltage, then select the filter function.

XCaution

To avoid possible damage to the Meter or to the equipment under test, follow these guidelines:

Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or capacitance.

Before measuring current, check the Meter's fuses. See " Testing the Fuses”.

3

Image 11
Contents Series Lifetime Limited Warranty Table of Contents Series List of Tables Series List of Figures Series Contacting Fluke IntroductionXWWarning Precautions and Safety InformationDigital Multimeter Electrical Symbols Electrical SymbolsSpecifications General SpecificationsModels 83 and 88 AC Voltage Function Specifications Detailed SpecificationsModel 87 AC Voltage Function Specifications Function Range Resolution Accuracy Model Current Function SpecificationsFrequency Counter Sensitivity and Trigger Levels Capacitance and Diode Function SpecificationsFrequency Counter Specifications Models 87 Frequency Counter Specifications ModelMIN MAX Recording Specifications Electrical Characteristics of the TerminalsOpening the Meter Case Basic MaintenanceCleaning the Meter Opening the Meter, Battery and Fuse Replacement Accessing the PCA and Replacing the LCDReassembling the Meter Case Removing LCD Mask to Access LCDTesting Fuses and Current Circuitry Replacing the BatteryReplacing the Fuses Required EquipmentPerformance Tests Trig Y Testing the FusesTesting the Display Basic Operability TestsCAL Accuracy Tests Testing the PushbuttonsTesting Meter Accuracy 600 Ω 330 Ω Use 2 wire Comp 328.5 to 329.1 to Ohms Press backlight button Backlight off Testing the Inductive Pickup 88 OnlyDegrees Setup for Inductive Pickup Test Waveform for Inductive Pickup Test Calibration Adjustment Password Calibration AdjustmentCalibration Adjustment Counter Restoring the Default Password Meter Buttons Used in the Calibration Steps Calibration Adjustment ProcedureMA / UA Series V Final Assembly Ref Des Description Part Qty Number Service and PartsSeries V Final Assembly