Fluke 80 Series V Testing the Inductive Pickup 88 Only, Press backlight button Backlight off

Page 27

Digital Multimeter

Performance Tests

Table 14. Accuracy Tests (cont.)

Step

Test

 

Range

5500A Output

Display Reading

 

 

 

Function

 

83 and 88

87

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

47

L

 

6 V dc

8 Vpp, 2 kHz Sq.

Max = 5.896 to 6.104

48

(87 and 88 only)

 

 

Wave, DC offset 2 V

Min = -1.898 to -2.102

 

 

 

 

 

 

 

 

 

Peak Min/Max

 

 

 

 

 

 

 

 

 

 

 

 

 

49

mL

 

 

0 °C

-1.0 to 1.0

-1.0 to 1.0

 

 

 

 

 

 

 

 

 

 

 

 

50

(87 and 88 only)

 

 

100 °C

98.0 to 102.0

98.0 to 102.0

 

 

Temperature 3

 

 

 

 

 

 

 

 

 

 

 

 

51

Backlight

 

 

Press backlight button

Backlight comes on

52

 

 

 

Press backlight button

Backlight Intensifies

53

 

 

 

Press backlight button

Backlight off

1.

Or short test leads and use REL to offset test lead resistance.

 

 

2.

Remove test leads from unit.

 

 

 

3.

To ensure accurate measurement, the Meter and thermocouple adapter must be at the same temperature. After

 

connecting the thermocouple adapter to the Meter allow for reading to stabalize before recording display reading.

4.

The Meter accuracy is not specified at this input signal frequency with Low-pass filter selected. The display reading

 

shown, check that the Low-pass filter is active and follows an expected roll-off curve.

 

5.

Use REL to compensate for internal Meter and lead capacitance. Test leads must be disconnected from the

 

calibrator before using REL.

 

 

 

 

6.

W 10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum;

 

> 10 A unspecified.

 

 

 

 

Testing the Inductive Pickup (88 Only)

To test the inductive pickup, a function generator output will simulate automobile spark plug signals on a loop of wire containing a 10 e resistor. The pickup will be clamped to the wire and output voltage from the pickup will be monitored by an oscilloscope.

Perform the following procedure to test the inductive pickup:

1.Solder a 10-inch piece of 14 AWG wire to one end of a 10 e 1 % resistor.

2.Connect the other end of the 10 e resistor to the terminal LOW output of the function generator. Place the other end of the 14-AWG wire to the HIGH output of the function generator. See Figure 5.

3.Connect a 10X scope probe from channel 2 (dc-coupled) of the oscilloscope across the 10 e resistor.

4.Clamp the inductive pickup to the wire loop on the HIGH side of the resistor as shown in Figure 5.

Make sure that the jaws of the inductive pickup are closed completely, and that the side of the inductive pickup that says "SPARK PLUG SIDE" points toward the HIGH output of the function generator.

5.Connect a 10X scope probe from channel 1 (dc-coupled) of the oscilloscope across the output of the inductive pickup.

19

Image 27
Contents Series Lifetime Limited Warranty Table of Contents Series List of Tables Series List of Figures Series Contacting Fluke IntroductionXWWarning Precautions and Safety InformationDigital Multimeter Electrical Symbols Electrical SymbolsSpecifications General SpecificationsDetailed Specifications Model 87 AC Voltage Function SpecificationsModels 83 and 88 AC Voltage Function Specifications Function Range Resolution Accuracy Model Current Function SpecificationsFrequency Counter Sensitivity and Trigger Levels Capacitance and Diode Function SpecificationsFrequency Counter Specifications Models 87 Frequency Counter Specifications ModelMIN MAX Recording Specifications Electrical Characteristics of the TerminalsBasic Maintenance Cleaning the MeterOpening the Meter Case Opening the Meter, Battery and Fuse Replacement Accessing the PCA and Replacing the LCDReassembling the Meter Case Removing LCD Mask to Access LCDTesting Fuses and Current Circuitry Replacing the BatteryReplacing the Fuses Required EquipmentPerformance Tests Trig Y Testing the FusesTesting the Display Basic Operability TestsCAL Testing the Pushbuttons Testing Meter AccuracyAccuracy Tests 600 Ω 330 Ω Use 2 wire Comp 328.5 to 329.1 to Ohms Press backlight button Backlight off Testing the Inductive Pickup 88 OnlyDegrees Setup for Inductive Pickup Test Waveform for Inductive Pickup Test Calibration Adjustment Calibration Adjustment CounterCalibration Adjustment Password Restoring the Default Password Meter Buttons Used in the Calibration Steps Calibration Adjustment ProcedureMA / UA Series V Final Assembly Ref Des Description Part Qty Number Service and PartsSeries V Final Assembly