Fluke 80 Series V manual Degrees

Page 28

80 Series V

Calibration Manual

6. Set up the function generator as follows:

Frequency

900 kHz

Amplitude

8.3 V

Burst Mode

yes

Burst Count

1

Burst Rate

100 Hz

Burst Phase

-90 Degrees

7.Set the oscilloscope for 0.5 V/DIV @ 0.5 ∝s/DIV.

8.Trigger the waveform on channel 2.

9.Adjust the amplitude of the function generator to produce a 3 VP-P triangle wave. See Figure 6.

10.Set the oscilloscope for 1.0 V/DIV @ 5.0 ms/DIV.

11.Trigger the waveform on Channel 1.

12.Check that the peak voltage is greater than 5.7 V and decays to less than 1.0 V between pulses. See Figure 6. Record the peak value for later use.

13.Adjust the function generator output so the peak voltage is 6 V.

14.Set the scope for 1.0 ms/DIV and trigger waveform.

15.Check the amplitude after 3.0 ms from the waveform peak, the voltage amplitude is 2.4 V +0.5/-0.8 (1.6 V to 2.9 V). (See Figure 3-5(C).)

16.Re-adjust the function generator output to obtain the value recorded in step 12.

17.Set the scope for 5.0 ms/DIV.

18.Turn the inductive pickup so that "SPARK PLUG SIDE" points along the wire connected to the LOW output of the function generator. Check that the waveform is less than 2 V.

20

Image 28
Contents Series Lifetime Limited Warranty Table of Contents Series List of Tables Series List of Figures Series Introduction Contacting FlukePrecautions and Safety Information XWWarningDigital Multimeter Electrical Symbols SpecificationsGeneral Specifications Electrical SymbolsModel 87 AC Voltage Function Specifications Detailed SpecificationsModels 83 and 88 AC Voltage Function Specifications Current Function Specifications Function Range Resolution Accuracy ModelCapacitance and Diode Function Specifications Frequency Counter Specifications Models 87Frequency Counter Specifications Model Frequency Counter Sensitivity and Trigger LevelsElectrical Characteristics of the Terminals MIN MAX Recording SpecificationsCleaning the Meter Basic MaintenanceOpening the Meter Case Accessing the PCA and Replacing the LCD Opening the Meter, Battery and Fuse ReplacementRemoving LCD Mask to Access LCD Reassembling the Meter CaseReplacing the Battery Testing Fuses and Current CircuitryRequired Equipment Replacing the FusesPerformance Tests Testing the Fuses Testing the DisplayBasic Operability Tests Trig YCAL Testing Meter Accuracy Testing the PushbuttonsAccuracy Tests 600 Ω 330 Ω Use 2 wire Comp 328.5 to 329.1 to Ohms Testing the Inductive Pickup 88 Only Press backlight button Backlight offDegrees Setup for Inductive Pickup Test Waveform for Inductive Pickup Test Calibration Adjustment Counter Calibration AdjustmentCalibration Adjustment Password Restoring the Default Password Calibration Adjustment Procedure Meter Buttons Used in the Calibration StepsMA / UA Service and Parts Series V Final Assembly Ref Des Description Part Qty NumberSeries V Final Assembly