Fluke 80 Series V manual Performance Tests

Page 22

80 Series V

Calibration Manual

Table 13. Required Equipment

Equipment

Required Characteristics

Recommended Model

 

 

 

Calibrator

AC Voltage Range: 0 - 1000 V ac

Fluke 5520A Multi-Product

 

Accuracy: ± 0.12 %

Calibrator or equivalent

 

Frequency Range: 60 - 20000 Hz

 

 

Accuracy: ± 3 %

 

 

DC Voltage Range: 0 - 1000 V dc

 

 

Accuracy: ± 0.012 %

 

 

Current Range: 350 A - 2 A

 

 

Accuracy: AC (60 Hz to 1 kHz): ± 0.25 %

 

 

DC: ± 0.05 %

 

 

Frequency Source: 19.999 kHz - 199.99 kHz

 

 

Accuracy: ± 0.0025 %

 

 

Amplitude: 150 mV to 6V rms

 

 

Accuracy: ± 5 %

 

 

Range: 1 - 100 M

 

 

Accuracy: 0.065 %

 

 

 

 

Function Generator

Frequency = 900 kHz

HP33120

 

Amplitude = 8.3V

 

 

Burst mode = 1

 

 

Burst rate = 100Hz

 

 

Burst Phase = -90 degrees

 

Fluke 80 AK

K-type

Fluke 80 AK

TC Adapter Accessory

 

 

 

 

 

K-type Thermocouple

K-type, mini-plug on both ends

 

 

 

 

Performance Tests

The following performance tests verify the complete operability of the Meter and check the accuracy of each Meter function against the Meter’s specifications. Performance tests should be performed annually to ensure that the Meter is within accuracy specifications.

Accuracy specifications are valid for a period of one year after calibration adjustment, when measured at an operating temperature of 18 °C to 28 °C and at a maximum of 90 % relative humidity.

To perform the following tests, it is not necessary to open the case. No adjustments are necessary. Make the required connections, apply the designated inputs, and determine if the reading on the Meter display falls within the acceptable range indicated.

Note

If the Meter fails any of these tests, it needs calibration adjustment or repair.

14

Image 22
Contents Series Lifetime Limited Warranty Table of Contents Series List of Tables Series List of Figures Series Introduction Contacting FlukePrecautions and Safety Information XWWarningDigital Multimeter General Specifications Electrical SymbolsSpecifications Electrical SymbolsModel 87 AC Voltage Function Specifications Detailed SpecificationsModels 83 and 88 AC Voltage Function Specifications Current Function Specifications Function Range Resolution Accuracy ModelFrequency Counter Specifications Model Capacitance and Diode Function SpecificationsFrequency Counter Specifications Models 87 Frequency Counter Sensitivity and Trigger LevelsElectrical Characteristics of the Terminals MIN MAX Recording SpecificationsCleaning the Meter Basic MaintenanceOpening the Meter Case Accessing the PCA and Replacing the LCD Opening the Meter, Battery and Fuse Replacement Removing LCD Mask to Access LCD Reassembling the Meter CaseReplacing the Battery Testing Fuses and Current CircuitryRequired Equipment Replacing the FusesPerformance Tests Basic Operability Tests Testing the FusesTesting the Display Trig YCAL Testing Meter Accuracy Testing the PushbuttonsAccuracy Tests 600 Ω 330 Ω Use 2 wire Comp 328.5 to 329.1 to Ohms Testing the Inductive Pickup 88 Only Press backlight button Backlight offDegrees Setup for Inductive Pickup Test Waveform for Inductive Pickup Test Calibration Adjustment Counter Calibration AdjustmentCalibration Adjustment Password Restoring the Default Password Calibration Adjustment Procedure Meter Buttons Used in the Calibration StepsMA / UA Service and Parts Series V Final Assembly Ref Des Description Part Qty NumberSeries V Final Assembly