80 Series V
Calibration Manual
Table 13. Required Equipment
Equipment | Required Characteristics | Recommended Model |
|
|
|
Calibrator | AC Voltage Range: 0 - 1000 V ac | Fluke 5520A |
| Accuracy: ± 0.12 % | Calibrator or equivalent |
| Frequency Range: 60 - 20000 Hz |
|
| Accuracy: ± 3 % |
|
| DC Voltage Range: 0 - 1000 V dc |
|
| Accuracy: ± 0.012 % |
|
| Current Range: 350 ∝A - 2 A |
|
| Accuracy: AC (60 Hz to 1 kHz): ± 0.25 % |
|
| DC: ± 0.05 % |
|
| Frequency Source: 19.999 kHz - 199.99 kHz |
|
| Accuracy: ± 0.0025 % |
|
| Amplitude: 150 mV to 6V rms |
|
| Accuracy: ± 5 % |
|
| Range: 1 Ω - 100 MΩ |
|
| Accuracy: 0.065 % |
|
|
|
|
Function Generator | Frequency = 900 kHz | HP33120 |
| Amplitude = 8.3V |
|
| Burst mode = 1 |
|
| Burst rate = 100Hz |
|
| Burst Phase = |
|
Fluke 80 AK | Fluke 80 AK | |
TC Adapter Accessory |
|
|
|
|
|
| ||
|
|
|
Performance Tests
The following performance tests verify the complete operability of the Meter and check the accuracy of each Meter function against the Meter’s specifications. Performance tests should be performed annually to ensure that the Meter is within accuracy specifications.
Accuracy specifications are valid for a period of one year after calibration adjustment, when measured at an operating temperature of 18 °C to 28 °C and at a maximum of 90 % relative humidity.
To perform the following tests, it is not necessary to open the case. No adjustments are necessary. Make the required connections, apply the designated inputs, and determine if the reading on the Meter display falls within the acceptable range indicated.
Note
If the Meter fails any of these tests, it needs calibration adjustment or repair.
14