Fluke 80 Series V manual Testing the Pushbuttons, Testing Meter Accuracy, Accuracy Tests

Page 25

Digital Multimeter

Performance Tests

Testing the Pushbuttons

To test the pushbuttons

1.Turn the Meter rotary knob to J.

2.Press each button and note that the meter responds with a beep for each button press.

3.Press and hold Ba second time to exit MIN MAX mode.

Testing Meter Accuracy

Perform the accuracy test steps in Table 15.

Table 15. Accuracy Tests

Step

Test

Range

5500A Output

Display Reading

 

Function

83 and 88

 

87

 

 

 

 

1

 

K

 

600 mV

330 mV, 60 Hz

327.9 to 332.1

327.3 to 332.7

2

 

 

600 mV

600 mV, 13 kHz

N/A

586.0 to 614.0

 

AC Volts

 

3

 

 

6 V

3.3 V, 60 Hz

3.281 to 3.319

3.275 to 3.325

 

 

 

4

 

 

 

6 V

3.3 V, 20 kHz

N/A

3.214 to 3.386

5

 

 

 

60 V

33 V, 60 Hz

32.81 to 33.19

32.75 to 33.25

6

 

 

 

60 V

33 V, 20 kHz

N/A

32.14 to 33.86

7

 

 

 

600 V

330 V, 60 Hz

328.1 to 331.9

327.5 to 332.5

8

 

 

 

600 V

330 V, 2.5 kHz

N/A

323.0 to 337.0

9

 

 

 

1000 V

500 V, 60 Hz

495 to 505

494 to 506

10

 

 

 

1000 V

1000 V, 1 kHz

986 to 1014

986 to 1014

11

 

K Hz

 

600 mV

150 mV, 99.95 kHz

99.93 to 99.97

99.93 to 99.97

12

 

 

600 mV

150 mV, 199.50 kHz

199.48 to 199.52

199.48 to 199.52

 

AC Volts

 

 

 

 

 

 

 

 

 

 

Frequency

 

 

 

 

 

13

 

Sensitivity

 

6 V

0.7 V, 99.95 kHz

99.93 to 99.97

99.93 to 99.97

14

 

 

 

60 V

7 V, 99.95 kHz

99.93 to 99.97

99.93 to 99.97

15

 

VHz

 

6 V

3.4 V, 1 kHz Sq. Wave

999.8 to 1000.2

999.8 to 1000.2

 

 

Trigger level

 

 

 

 

 

16

 

VHz

 

6 V

5 V, 1 kHz, DC offset 2.5 V

49.7% to 50.3 %

49.7 % to 50.3 %

 

 

Duty Cycle

 

 

Sq. Wave

 

 

 

 

 

 

 

 

 

17

 

V

 

6V

3.3 V dc

3.296 to 3.304

3.297 to 3.303

18

 

 

60 V

33 V dc

32.96 to 33.04

32.97 to 33.03

 

DC Volts

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

19

 

 

 

600 V

330 V dc

329.6 to 330.4

329.7 to 330.3

 

 

 

 

 

 

 

 

20

 

 

 

1000 V

1000 V dc

998 to 1002

998 to 1002

 

 

 

 

 

 

 

 

21

 

mV

 

600 mV

33 mV dc

32.8 to 33.2

32.9 to 33.1

22

 

DC Volts

 

600 mV

330 mV dc

328.9 to 331.1

329.6 to 330.4

17

Image 25
Contents Series Lifetime Limited Warranty Table of Contents Series List of Tables Series List of Figures Series Contacting Fluke IntroductionXWWarning Precautions and Safety InformationDigital Multimeter Specifications Electrical SymbolsGeneral Specifications Electrical SymbolsModel 87 AC Voltage Function Specifications Detailed SpecificationsModels 83 and 88 AC Voltage Function Specifications Function Range Resolution Accuracy Model Current Function SpecificationsFrequency Counter Specifications Models 87 Capacitance and Diode Function SpecificationsFrequency Counter Specifications Model Frequency Counter Sensitivity and Trigger LevelsMIN MAX Recording Specifications Electrical Characteristics of the TerminalsCleaning the Meter Basic MaintenanceOpening the Meter Case Opening the Meter, Battery and Fuse Replacement Accessing the PCA and Replacing the LCDReassembling the Meter Case Removing LCD Mask to Access LCDTesting Fuses and Current Circuitry Replacing the BatteryReplacing the Fuses Required EquipmentPerformance Tests Testing the Display Testing the FusesBasic Operability Tests Trig YCAL Testing Meter Accuracy Testing the PushbuttonsAccuracy Tests 600 Ω 330 Ω Use 2 wire Comp 328.5 to 329.1 to Ohms Press backlight button Backlight off Testing the Inductive Pickup 88 OnlyDegrees Setup for Inductive Pickup Test Waveform for Inductive Pickup Test Calibration Adjustment Counter Calibration AdjustmentCalibration Adjustment Password Restoring the Default Password Meter Buttons Used in the Calibration Steps Calibration Adjustment ProcedureMA / UA Series V Final Assembly Ref Des Description Part Qty Number Service and PartsSeries V Final Assembly