Fluke 80 Series V manual Capacitance and Diode Function Specifications

Page 15

Digital Multimeter

Specifications

Table 7. Capacitance and Diode Function Specifications

Function

Range

Resolution

Accuracy

 

 

 

 

E

10.00 nF

0.01 nF

± (1 % + 2)1

 

100.0 nF

0. 1 nF

± (1 % + 2)1

 

1.000

F

0.001 F

± (1 % + 2)

 

10.00

F

0.01 F

± (1 % + 2)

 

100.0

F

0.1 F

± (1 % + 2)

 

9999 F

1 F

± (1 % + 2)

 

 

 

 

 

G

3.000

V

0.001 V

± (2 % + 1)

 

 

 

 

 

1. With a film capacitor or better, using Relative mode to zero residual.

Table 8. Frequency Counter Specifications (Models 87 and 83)

Function

Range

Resolution

Accuracy

 

 

 

 

Frequency

199.99

0.01 Hz

± (0.005 % + 1)

(0.5 Hz to 200 kHz,

1999.9

0.1 Hz

± (0.005 % + 1)

pulse width > 2 s)

19.999 kHz

0.001 kHz

± (0.005 % + 1)

 

199.99 kHz

0.01 kHz

± (0.005 % + 1)

 

> 200 kHz

0.1 kHz

unspecified

 

 

 

 

Table 9. Frequency Counter Specifications (Model 88)

 

Function

Range

Resolution

Accuracy

Pulse Width

Resolution

 

Range (ms)1

(ms)

 

 

 

 

 

Frequency2

199.99

0.01 Hz

± (0.01 % + 1)

1999.9

0.1

(0.5 Hz to 200 kHz,

1999.9

0.1 Hz

± (0.01 % + 1)

5.00

0.01

pulse width > 2 s)

19.999 kHz

0.001 kHz

± (0.01 % + 1)

0.500

0.001

 

 

199.99 kHz

0.01 kHz

± (0.01 % + 1)

0.0500

0.0001

 

 

> 200 kHz

0.1 kHz

unspecified

 

 

 

 

 

 

 

 

RPM n

30 to 9,000

1 RPM

± 2 RPM

 

 

 

 

 

 

 

 

RPM o

60 to 12,000

1 RPM

± 2 RPM

 

 

 

 

 

 

 

 

% Duty Cycle3

0.0 to 99.9% (0.5 Hz to 200 kHz, Pulse Width >2 s)

 

 

Pulse Width3

0.002 to 1999.9 ms (4 Hz to 200 kHz, Pulse Width >2 s)

 

 

1.

Pulse Width range is determined by the frequency of the signal.

 

 

2.

Frequency measurements can be made on voltage or current inputs. The current inputs are always dc coupled.

3. For rise times >1 s. Duty Cycle Accuracy: ±(0.2 % per kHz + 0.1%). Pulse Width Accuracy: ±(0.002 ms + 3).

Table 10. Frequency Counter Sensitivity and Trigger Levels

Input Range1

Minimum Sensitivity (RMS Sine wave)

 

Approximate Trigger Level

 

 

 

(DC Voltage Function)

5 Hz - 20 kHz

0.5 Hz - 200 kHz

 

 

 

 

 

 

600 mV dc

70 mV (to 400 Hz)

70 mV (to 400 Hz)

 

40 mV

600 mV ac

150 mV

150 mV

 

6 V

0.3 V

0.7 V

 

1.7 V

60 V

3 V

7 V (140 kHz)

 

4 V

600 V

30 V

70 V (14.0 kHz)

 

40 V

1000 V

100 V

200 V (1.4 kHz)

 

100 V

 

 

 

 

 

Duty Cycle Range

 

 

Accuracy

 

 

0.0 to 99.9 %

Within ± (0.2% per kHz + 0.1 %) for rise times < 1 s.

 

 

 

 

 

1. Maximum input for specified accuracy = 10X Range or 1000 V.

7

Image 15
Contents Series Lifetime Limited Warranty Table of Contents Series List of Tables Series List of Figures Series Contacting Fluke IntroductionXWWarning Precautions and Safety InformationDigital Multimeter Electrical Symbols Electrical SymbolsSpecifications General SpecificationsDetailed Specifications Model 87 AC Voltage Function SpecificationsModels 83 and 88 AC Voltage Function Specifications Function Range Resolution Accuracy Model Current Function SpecificationsFrequency Counter Sensitivity and Trigger Levels Capacitance and Diode Function SpecificationsFrequency Counter Specifications Models 87 Frequency Counter Specifications ModelMIN MAX Recording Specifications Electrical Characteristics of the TerminalsBasic Maintenance Cleaning the MeterOpening the Meter Case Opening the Meter, Battery and Fuse Replacement Accessing the PCA and Replacing the LCDReassembling the Meter Case Removing LCD Mask to Access LCDTesting Fuses and Current Circuitry Replacing the BatteryReplacing the Fuses Required EquipmentPerformance Tests Trig Y Testing the FusesTesting the Display Basic Operability TestsCAL Testing the Pushbuttons Testing Meter AccuracyAccuracy Tests 600 Ω 330 Ω Use 2 wire Comp 328.5 to 329.1 to Ohms Press backlight button Backlight off Testing the Inductive Pickup 88 OnlyDegrees Setup for Inductive Pickup Test Waveform for Inductive Pickup Test Calibration Adjustment Calibration Adjustment CounterCalibration Adjustment Password Restoring the Default Password Meter Buttons Used in the Calibration Steps Calibration Adjustment ProcedureMA / UA Series V Final Assembly Ref Des Description Part Qty Number Service and PartsSeries V Final Assembly