Digital Multimeter
Specifications
664 to 600 V Overvoltage Category IV. UL listed to
Detailed Specifications
For all detailed specifications:
Accuracy is given as ±([% of reading] + [number of least significant digits]) at 18° C to 28° C, with relative humidity up to 90 %, for a period of one year after calibration.
For Model 87 in the 4
Table 2. Model 87 AC Voltage Function Specifications
Function | Range | Resolution | Accuracy |
|
|
|
|
|
|
|
|
|
| 45 - 65 Hz | 30 - 200 Hz | 200 - 440 Hz | 440 Hz - | 1 - 5 kHz |
| 5 - 20 kHz1 | ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 1 kHz |
|
|
|
|
| K2,4 |
| 600.0 mV | 0.1 mV | ± (0.7 % + 4) |
|
|
|
|
| ± (2.0 % + 4) |
| ± (2.0 % + 20) | |||||||
|
|
| 6.000 V | 0.001 V |
|
|
|
|
|
|
|
|
|
|
| |||||
|
|
| 60.00 V | 0.01 V | ± (0.7 % + 2) | ± (1.0 % + 4) |
|
|
|
|
| |||||||||
|
|
| 600.0 V | 0.1 V |
|
|
|
|
|
|
|
|
|
| ± (2.0 % + 4)3 |
| unspecified | |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| 1000 V | 1 V |
|
|
|
|
|
|
|
|
|
| unspecified |
| unspecified | |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Low pass filter |
|
|
|
|
| ± (0.7 % + 2) | ± (1.0 % + 4) |
| +1 % + 4 | unspecified | unspecified |
| unspecified | ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
1. | Below 10 % of range, add 12 counts. |
|
|
|
|
|
|
|
|
| ||||||||||
2. | The Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter | |||||||||||||||||||
|
| may display a residual reading between 1 and 30 counts. A 30 count residual reading will cause only a | ||||||||||||||||||
|
| change for readings over 3 % of range. Using REL to offset this reading may produce a much larger constant error | ||||||||||||||||||
|
| in later measurements. |
|
|
|
|
|
|
|
|
|
|
| |||||||
3. | Frequency range: 1 kHz to 2.5 kHz. |
|
|
|
|
|
|
|
|
| ||||||||||
4. | A residual reading of up to 13 digits with leads shorted, will not affect stated accuracy above 3 % of range. | |||||||||||||||||||
5. | Specification increases from |
|
|
|
| |||||||||||||||
|
|
|
| Table 3. Models 83 and 88 AC Voltage Function Specifications |
| |||||||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
| Function | Range |
|
| Resolution |
|
|
|
|
| Accuracy |
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
| 50 Hz - 60 Hz |
| 30 Hz - 1 kHz |
| 1 kHz - 5 kHz | ||||
| K1 |
|
| 600.0 mV |
| 0.1 mV |
|
| ± (0.5 % + 4) |
| ± (1.0 % + 4) |
| ± (2.0 % + 4) | |||||||
|
|
|
| 6.000 V |
|
| 0.001 V |
| ± (0.5 % + 2) |
| ± (1.0 % + 4) |
| ± (2.0 % + 4) | |||||||
|
|
|
| 60.00 V |
|
| 0.01 V |
|
| ± (0.5 % + 2) |
| ± (1.0 % + 4) |
| ± (2.0 % + 4) | ||||||
|
|
|
| 600.0 V |
|
| 0.1 V |
|
| ± (0.5 % + 2) |
| ± (1.0 % + 4) |
| ± (2.0 % + 4)2 | ||||||
|
|
|
| 1000 V |
|
| 1 V |
|
| ± (0.5 % + 2) |
| ± (1.0 % + 4) |
| unspecified | ||||||
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||
| 1. | Below a reading of 200 counts, add 10 counts. |
|
|
|
|
|
|
| |||||||||||
| 2. | Frequency range: | 1 kHz to 2.5 kHz. |
|
|
|
|
|
|
|
|
| ||||||||
|
| For models 83 and 88, ac conversions are |
|
5