Fluke 80 Series V manual Current Function Specifications, Function Range Resolution Accuracy Model

Page 14

80 Series V

Calibration Manual

Table 4. DC Voltage, Resistance, and Conductance Function Specifications

Function

Range

Resolution

 

Accuracy

 

Model 83

Model 87

Model 88

 

 

 

 

 

 

 

 

 

 

 

L

6.000

V

0.001 V

± (0.1 % + 1)

± (0.05 % + 1)

± (0.1 % + 1)

60.00

V

0.01 V

± (0.1 % + 1)

± (0.05 % + 1)

± (0.1 % + 1)

 

600.0

V

0.1 V

± (0.1 % + 1)

± (0.05 % + 1)

± (0.1 % + 1)

 

1000 V

1 V

± (0.1 % + 1)

± (0.05 % + 1)

± (0.1 % + 1)

 

 

 

 

 

 

 

F

600.0 mV

0.1 mV

± (0.3 % + 1)

± (0.1 % + 1)

± (0.3 % + 1)

mV

e

600.0

0.1

± (0.4 % + 2)1

± (0.2 % + 2)1

± (0.4 % + 2)1

6.000 k

0.001 k

± (0.4 % + 1)

± (0.2 % + 1)

± (0.4 % + 1)

 

60.00 k

0.01 k

± (0.4 % + 1)

± (0.2 % + 1)

± (0.4 % + 1)

 

600.0 k

0.1 k

± (0.7 % + 1)

± (0.6 % + 1)

± (0.7 % + 1)

 

6.000

M

0.001 M

± (0.7 % + 1)

± (0.6 % + 1)

± (0.7 % + 1)

nS

50.00

M

0.01 M

± (1.0 % + 3)2

± (1.0 % + 3)2

± (1.0 % + 3)2

60.00 nS

0.01 nS

± (1.0 % + 10)1

± (1.0 % + 10)1

± (1.0 % + 10)1

1. When using the REL function to compensate for offsets.

2. Add 0.5 % of reading when measuring above 30 Min the 50 Mrange, and 20 counts below 33 nS in the 60 nS range.

Table 5. Temperature Specifications (87 and 88 Only)

 

 

Temperature

 

Resolution

Accuracy1,2

-200

°C to +1090 °C

0.1

°C

1 % + 10

-328

°F to +1994 °F

0.1

°F

1 % + 18

1.

Does not include error of the thermocouple probe.

 

2.

Accuracy specification assumes ambient temperature stable to ± 1 °C. For ambient temperature changes of ± 5 °C,

rated accuracy applies after 1 hour.

Table 6. Current Function Specifications

Function

Range

Resolution

 

Accuracy

 

Burden

Model 831

Model 872, 3

Model 881

Voltage

 

 

 

 

 

 

 

 

(typical)

mA

 

60.00 mA

0.01 mA

± (1.2 % + 2)5

± (1.0 % + 2)

± (1.2 % + 2)5

1.8 mV/mA

\

400.0 mA6

0.1 mA

± (1.2 % + 2)5

± (1.0 % + 2)

± (1.2 % + 2)5

1.8 mV/mA

6.000

A

0.001 A

± (1.2 % + 2)5

± (1.0 % + 2)

± (1.2 % + 2)5

0.03 V/A

(45 Hz to

10.00

A4

0.01 A

± (1.2 % + 2)5

± (1.0 % + 2)

± (1.2 % + 2)5

0.03 V/A

2 kHz)

 

 

 

 

 

 

 

 

 

60.00 mA

0.01 mA

± (0.4 % + 4)

± (0.2 % + 4)

± (0.4 % + 4)

1.8 mV/mA

mA

 

400.0 mA6

0.1 mA

± (0.4 % + 2)

± (0.2 % + 2)

± (0.4 % + 2)

1.8 mV/mA

[

6.000

A

0.001 A

± (0.4 % + 4)

± (0.2 % + 4)

± (0.4 % + 4)

0.03 V/A

A B

10.00

A4

0.01 A

± (0.4 % + 2)

± (0.2 % + 2)

± (0.4 % + 2)

0.03 V/A

600.0

A

0.1 A

± (1.2 % + 2)5

± (1.0 % + 2)

± (1.2 % + 2)5

100 V/A

(45 Hz to

6000 A

1 A

± (1.2 % + 2)5

± (1.0 % + 2)

± (1.2 % + 2)5

100 V/A

2 kHz)

 

 

 

 

 

 

 

AF

600.0

A

0.1 A

± (0.4 % + 4)

± (0.2 % + 4)

± (0.4 % + 4)

100 V/A

6000 A

1 A

± (0.4 % + 2)

± (0.2 % + 2)

± (0.4 % + 2)

100 V/A

 

 

 

 

 

 

 

 

1.

AC conversion for Model 83 and 88 is ac coupled and calibrated to the rms value of a sine wave input.

2.

AC conversions for Model 87 are ac coupled, true rms responding, and valid from 3 % to 100 % of range, except

 

400 mA range (5 % to 100 % of range) and 10 A range (15 % to 100 % or range).

 

3.

Model 87 is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter

 

may display a residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2 digit

 

change for readings over 3 % of range. Using REL to offset this reading may produce a much larger constant

 

error in later measurements

 

 

 

 

4.

W 10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum;

 

> 10 A unspecified.

 

 

 

 

 

5.

Below a reading of 200 counts, add 10 counts.

 

 

 

6.

400 mA continuous; 600 mA for 18 hrs maximum.

 

 

 

6

Image 14
Contents Series Lifetime Limited Warranty Table of Contents Series List of Tables Series List of Figures Series Introduction Contacting FlukePrecautions and Safety Information XWWarning Digital Multimeter General Specifications Electrical SymbolsSpecifications Electrical SymbolsModels 83 and 88 AC Voltage Function Specifications Detailed SpecificationsModel 87 AC Voltage Function Specifications Current Function Specifications Function Range Resolution Accuracy ModelFrequency Counter Specifications Model Capacitance and Diode Function SpecificationsFrequency Counter Specifications Models 87 Frequency Counter Sensitivity and Trigger LevelsElectrical Characteristics of the Terminals MIN MAX Recording SpecificationsOpening the Meter Case Basic MaintenanceCleaning the Meter Accessing the PCA and Replacing the LCD Opening the Meter, Battery and Fuse ReplacementRemoving LCD Mask to Access LCD Reassembling the Meter CaseReplacing the Battery Testing Fuses and Current CircuitryRequired Equipment Replacing the FusesPerformance Tests Basic Operability Tests Testing the FusesTesting the Display Trig YCAL Accuracy Tests Testing the PushbuttonsTesting Meter Accuracy 600 Ω 330 Ω Use 2 wire Comp 328.5 to 329.1 to Ohms Testing the Inductive Pickup 88 Only Press backlight button Backlight offDegrees Setup for Inductive Pickup Test Waveform for Inductive Pickup Test Calibration Adjustment Password Calibration AdjustmentCalibration Adjustment Counter Restoring the Default Password Calibration Adjustment Procedure Meter Buttons Used in the Calibration StepsMA / UA Service and Parts Series V Final Assembly Ref Des Description Part Qty NumberSeries V Final Assembly