Teledyne manual T201 Analyzer Specifications, T201 Operating Specifications

Page 15

Teledyne API Model T201 NH3 Analyzer Operator Manual

T201 Ammonia Analyzer

1.4T201 ANALYZER SPECIFICATIONS

Table 1-1. T201 Operating Specifications

Ranges

Measurement Units

Noise at Zero1

Noise at Span1

Lower Detectable Limit2

Zero Drift3, 6

Span Drift6

Lag Time

Rise Time4

Fall Time4

Sample Flow Rate

Linearity

Temp Range

Dimensions H x W x D

Weight, Analyzer

Weight, Converter

Weight, Pump

Power, Analyzer

Power, Analyzer5

Power, Pump

Power, Pump CE Mark5

Environmental

Analog Output Resolution

standard

Serial I/O

optional

Min: 0-50 ppb Full scale

Max: 0-2000 ppb Full scale (selectable, independent NH3, NO, NO2, NOx ranges supported)

PPB, PPM, microgram/m3, milligram/m3 (user selectable)

<0.5 ppb RMS

<1.0% of reading above 50 ppb

1 ppb RMS

2 ppb / 24 hours

<1.0% FS Range / 24 hours

40 seconds

90% 300 seconds

90% 300 seconds

1000 cm3/min + 10% (500 cm3/min bypass to vacuum manifold, 500 cm3/min to reaction cell)

NO calibration 1% of full scale NH3 calibration 2% of full scale 15-40 oC

7” x 17” x 23.6” (18cm x 43cm x 61cm)

43 lbs (20 kg)

24 lbs (11 kg)

16 lbs (7 kg)

100V ~50/60 Hz, 120V ~60 Hz, 220V ~50Hz, 240V ~50 Hz, 125 watts 230V ~50 Hz, 125 watts

110V ~60 Hz, 220V ~50 Hz, 240V ~50 Hz, 295 watts 230 V ~50 Hz, 2.5 A peak

Installation Category Pollution Degree 2, Over-voltage Category II 1 part in 4096 of selected full-scale voltage (12 bit)

(1)Ethernet: 10/100Base-T; (2) RS-232 (300 – 115,200 baud)

(2)USB device ports; (8) opto-isolated digital outputs; (6) opto-isolated digital inputs; (4) analog outputs

(1)USB com port; (1) RS485 (300 – 115,200 baud); (8) analog inputs (0-10V, 12-bit); (4) digital alarm outputs; Multidrop RS232; (3) 4-20mA current outputs

1As defined by USEPA

2Defined as twice the zero noise level

3At constant temperature and voltage

4When pneumatics are conditioned with NH3 overnight and the gases (zero air to NH3) are manually switched at the sample inlet at the rear of the M501.

5Electrical rating for CE Mark Compliance

6Applies when sampling NH3; better results expected for NO/NOx gas measurements.

07271B DCN6646

13

Image 15
Contents Manual Addendum For use with the T200 Operators Manual, P/NSAN DIEGO, CA Page About this Manual This page intentionally left blank Safety Messages Mise EN Garde Consignes DE SécuritéTable of Contents 07271B DCN6646 Do not pressurize the Sample/Span/Zero port above ambient T201 Ammonia AnalyzerTeledyne API Model T201 NH3 Analyzer Operator Manual T201 and M501 NH 3 Pneumatic Flow Outlined in the following equations Theory of OperationMinimizing PMT Drift Sample Filtration Special Considerations for Ammonia MeasurementT201 Operating Specifications T201 Analyzer SpecificationsThis page intentionally left blank Calibration Procedure Step No Action Comment Zeroing the AnalyzerSpanning the Analyzer with Nitric Oxide GAS No Conv Spanning the Analyzer with Ammonia GAS Efficiency value through the user interface, if required TNx Conv Exit NH3 Calibration Procedure NH3 Gas through the Sample Port NH3 Calibration Procedure NH3 Gas through the Span Port This page intentionally left blank + + Configurable Analog Output OverviewStrip Chart HyperlinkAnalog Output Pin-Outs M501 NH3 Maintenance Maintenance ScheduleDescription Replacing the Catalytic CartridgeThermocouple Location Replacing the ThermocoupleTeledyne API Model T201 NH3 Analyzer Operator Manual Flow Check Autozero Flow CheckThis page intentionally left blank Alarm Alarms and CautionsAlarms and Cautions Tools ProcedureCalculating Moly Converter Efficiencies PurposeCalculating Moly Converter Efficiencies Moly Test Data Sheet This page intentionally left blank To our Website or contact Sales for recent updates T201 Spare Parts ListPCA,TEC DRIVER,PMT,KB Assy & TEST, Spare PS38 Teledyne API Model T201 NH3 Analyzer Operator ManualThis page intentionally left blank