Teledyne T201 manual Spanning the Analyzer with Ammonia GAS

Page 21

Teledyne API Model T201 NH3 Analyzer Operator Manual

Calibration Procedure

 

 

Table 2-4. NO Calibration Procedure - NO Gas through the SPAN Port

 

 

 

 

 

 

 

Step No.

Action

 

Comment

 

 

 

 

1.

Press CALS

The T201 enters the calibrate mode from sample

 

 

 

mode. The NO gas is supplied through the SPAN inlet

 

 

 

port on the rear panel of the M501 NH3.

 

2

Table 2.3

Follow Steps 2 to 5 in Table 2-3.

 

 

 

 

Notes

Both the TNx and NOx channels must be spanned individually with nitric

 

 

oxide gas. When you span the NOx

channel both the NO and NOx

concentrations must be manually inputted.

It is recommended that the calibration of both the TNx and NOx channels be done at the same time.

Always confirm that SPAN gas is flowing past the sample or span port on the rear of the M501 NH3 converter. There should always be some excess flow. If insufficient flow is supplied to the analyzer ambient gas will be entrained and the span values will be incorrect. However, it is important not to over pressurize the analyzer.

It is a good idea to allow the analyzer to sample span gas for extended periods following span calibration. Large drifts can signify leaks or indicate an issue with the span gas source.

If after spanning with nitric oxide, the analyzer slopes are out of range conduct the following:

-Confirm the gas sources are good. Usually the bottle is certified to a known level of nitric oxide. The bottle should be balanced with nitrogen.

-Try not to use the same regulator for nitric oxide as was used on ammonia. This also holds for the tubing between the different bottles and the mixing source (calibrator).

-Confirm the zero reading is good by delivering Zero gas.

-Confirm that the concentration delivered to the analyzer is equal to what you entered under the CONC menu NOx and TNx.

-Leak check both the M501 NH3 and analyzer together.

-Using section 11.6.5 of the T200 manual confirm that the PMT sensor hardware calibration is set correctly.

2.3SPANNING THE ANALYZER WITH AMMONIA GAS

07271B DCN6646

19

Image 21
Contents Manual Addendum For use with the T200 Operators Manual, P/NSAN DIEGO, CA Page About this Manual This page intentionally left blank Safety Messages Mise EN Garde Consignes DE SécuritéTable of Contents 07271B DCN6646 Do not pressurize the Sample/Span/Zero port above ambient T201 Ammonia AnalyzerTeledyne API Model T201 NH3 Analyzer Operator Manual T201 and M501 NH 3 Pneumatic Flow Outlined in the following equations Theory of OperationMinimizing PMT Drift Sample Filtration Special Considerations for Ammonia MeasurementT201 Operating Specifications T201 Analyzer SpecificationsThis page intentionally left blank Calibration Procedure Step No Action Comment Zeroing the AnalyzerSpanning the Analyzer with Nitric Oxide GAS No Conv Spanning the Analyzer with Ammonia GAS Efficiency value through the user interface, if required TNx Conv Exit NH3 Calibration Procedure NH3 Gas through the Sample Port NH3 Calibration Procedure NH3 Gas through the Span Port This page intentionally left blank Strip Chart Configurable Analog Output OverviewHyperlink + +Analog Output Pin-Outs M501 NH3 Maintenance Maintenance ScheduleDescription Replacing the Catalytic CartridgeThermocouple Location Replacing the ThermocoupleTeledyne API Model T201 NH3 Analyzer Operator Manual Flow Check Autozero Flow CheckThis page intentionally left blank Alarm Alarms and CautionsAlarms and Cautions Calculating Moly Converter Efficiencies ProcedurePurpose ToolsCalculating Moly Converter Efficiencies Moly Test Data Sheet This page intentionally left blank To our Website or contact Sales for recent updates T201 Spare Parts ListPCA,TEC DRIVER,PMT,KB Assy & TEST, Spare PS38 Teledyne API Model T201 NH3 Analyzer Operator ManualThis page intentionally left blank