Page 17

2.0 CALIBRATION PROCEDURE

This section begins with a high-level overview of the calibration procedure for the T201 analyzer. Details are provided starting in Section 2.1.

First, take the time to read the T200 manual to familiarize yourself with the Chemiluminescence process. It is important to remember that the analyzer is merely measuring different levels of nitric oxide sample gas on three separate channels (TNx, NOx, NO). The ammonia and nitrogen dioxide concentrations are then calculated using this information.

Assemble the T201 analyzer according to the pneumatic configurations outlined in Figure 1-1.Remember: DO NOT pressurize the sample, span or zero ports (Zero/SPAN valves are optional) during calibration. Allow the analyzer to pull the gas through the system using the vacuum pump.

CAUTION!

If the presence of ozone is detected at any time, call Teledyne API Customer Service as soon as possible:

800-324-5190 or email: api-customerservice@teledyne.com

Next, zero the analyzer using an approved zero air source, such as, bottled zero air, nitrogen or zero air generated by a Teledyne-API M701 zero air source. This sets the zero offset for the three individual channels (TNx_OFFS, NOx_OFFS, NO_OFFS). Confirm that all displayed concentrations read zero.

Span the analyzer using bottled nitric oxide gas diluted to a level set to 80% of the expected sample range, preferably using a M702SS Calibrator. This sets the slope for the three individual channels (TNx_SLOPE, NOx_SLOPE, NO_SLOPE). Confirm that displayed concentrations TNx , NOx and NO display the correct span concentration.

If you haven’t done so recently, conduct a GPT (gas phase titration) and confirm the molybdenum is functioning accordingly. Section 8.3.3 of the T200 manual outlines the GPT procedure. Section 6.0 of this manual consists of a service note for checking the molybdenum converter. There should be no reason to adjust the molybdenum efficiency factor on a new analyzer. If a diluted bottled of nitrogen dioxide gas is used to determine the molybdenum efficiency, allow enough time for the span value to stabilize. Nitrogen dioxide exhibits similar hold up issues as ammonia gas. Therefore, it may take a number of hours before the NOx reading stabilizes.

Span the analyzer using bottled ammonia gas diluted to a level set to 80% of the expected sample range, preferably using a M702SS Calibrator.

Note

The first time bottled ammonia is connected to the gas dilution system,

 

regulator/calibrator, the TNx reading may take a number of hours to stabilize.

 

 

07271B DCN6646

15

Image 17
Contents SAN DIEGO, CA Manual AddendumFor use with the T200 Operators Manual, P/N Page About this Manual This page intentionally left blank Safety Messages Mise EN Garde Consignes DE SécuritéTable of Contents 07271B DCN6646 Do not pressurize the Sample/Span/Zero port above ambient T201 Ammonia AnalyzerTeledyne API Model T201 NH3 Analyzer Operator Manual T201 and M501 NH 3 Pneumatic Flow Outlined in the following equations Theory of OperationMinimizing PMT Drift Sample Filtration Special Considerations for Ammonia MeasurementT201 Operating Specifications T201 Analyzer SpecificationsThis page intentionally left blank Calibration Procedure Step No Action Comment Zeroing the AnalyzerSpanning the Analyzer with Nitric Oxide GAS No Conv Spanning the Analyzer with Ammonia GAS Efficiency value through the user interface, if required TNx Conv Exit NH3 Calibration Procedure NH3 Gas through the Sample Port NH3 Calibration Procedure NH3 Gas through the Span Port This page intentionally left blank Strip Chart Configurable Analog Output OverviewHyperlink + +Analog Output Pin-Outs M501 NH3 Maintenance Maintenance ScheduleDescription Replacing the Catalytic CartridgeThermocouple Location Replacing the ThermocoupleTeledyne API Model T201 NH3 Analyzer Operator Manual Flow Check Autozero Flow CheckThis page intentionally left blank Alarm Alarms and CautionsAlarms and Cautions Calculating Moly Converter Efficiencies ProcedurePurpose ToolsCalculating Moly Converter Efficiencies Moly Test Data Sheet This page intentionally left blank To our Website or contact Sales for recent updates T201 Spare Parts ListPCA,TEC DRIVER,PMT,KB Assy & TEST, Spare PS38 Teledyne API Model T201 NH3 Analyzer Operator ManualThis page intentionally left blank