Maintenance Schedule

Teledyne API Model T201 NH3 Analyzer Operator Manual

Once you have the right tools and parts, replace the thermocouple as follows:

1.Ensure power is removed from the M501 NH3 Converter. If the Converter has been operational you will need to wait for at least 2 hours for the Converter oven to cool before continuing with the replacement of the thermocouple.

2.Remove the cover from the Converter chassis.

3.Unscrew the (4) nuts that secure the front panel to the chassis. They are located just behind the front panel along the top.

4.Lower the front panel to gain easier access to the end of the quartz tube.

5.Unscrew the (3) nuts that secure the inner cover protecting the Heater Block and quartz tube. Remove this cover.

6.Cut the tie-wrap hold-down that secures the thermocouple to the fitting at the end of the quartz tube.

7.Loosen the Teflon fitting at the end of the quartz tube taking care not to put any stress on the tube, and slide the fitting off the tube.

8.Remove the thermocouple.

9.Disconnect the thermocouple wires from the Temperature Controller noting that the ‘Yellow’ wire goes to pin 1 and the ‘Red’ wire goes to pin 2.

10.The thermocouple should slide into the Heater Block.

11.Replace the tie-wrap hold-down.

12.Form the thermocouple wire so that it rests in the cavity with little movement.

13.Connect the (2) wires of the thermocouple to the Temperature Controller. The “yellow” wire should be connected to pin 1 and the “red” wire should be connected to pin 2.

14.At this point, all connections have been made, both electrically and pneumatically. A leak check should be performed on the Converter to verify that all connections are leak free. If a leak is detected, the leak should be resolved before continuing.

15.Install the inner cover of the Heater Block and secure with the (3) nuts. Close the front fanel and secure with the (4) nuts. Install the top cover on the Converter chassis.

16.The Converter is now ready for the application of power. You will be looking for an indication from the Temperature Controller that it is functioning correctly and driving the heater to the desired “set” temperature. Apply power now.

17.After the Converter comes to the regulated temperature, perform the Auto Tune function to tune the Temperature Controller to the new thermocouple.

30

07271B DCN6646

Image 32
Contents SAN DIEGO, CA Manual AddendumFor use with the T200 Operators Manual, P/N Page About this Manual This page intentionally left blank Safety Messages Consignes DE Sécurité Mise EN GardeTable of Contents 07271B DCN6646 T201 Ammonia Analyzer Do not pressurize the Sample/Span/Zero port above ambientTeledyne API Model T201 NH3 Analyzer Operator Manual T201 and M501 NH 3 Pneumatic Flow Theory of Operation Outlined in the following equationsMinimizing PMT Drift Special Considerations for Ammonia Measurement Sample FiltrationT201 Analyzer Specifications T201 Operating SpecificationsThis page intentionally left blank Calibration Procedure Zeroing the Analyzer Step No Action CommentSpanning the Analyzer with Nitric Oxide GAS No Conv Spanning the Analyzer with Ammonia GAS Efficiency value through the user interface, if required TNx Conv Exit NH3 Calibration Procedure NH3 Gas through the Sample Port NH3 Calibration Procedure NH3 Gas through the Span Port This page intentionally left blank Configurable Analog Output Overview Strip ChartHyperlink + +Analog Output Pin-Outs Maintenance Schedule M501 NH3 MaintenanceReplacing the Catalytic Cartridge DescriptionReplacing the Thermocouple Thermocouple LocationTeledyne API Model T201 NH3 Analyzer Operator Manual Autozero Flow Check Flow CheckThis page intentionally left blank Alarms and Cautions AlarmAlarms and Cautions Procedure Calculating Moly Converter EfficienciesPurpose ToolsCalculating Moly Converter Efficiencies Moly Test Data Sheet This page intentionally left blank T201 Spare Parts List To our Website or contact Sales for recent updatesPCA,TEC DRIVER,PMT,KB Teledyne API Model T201 NH3 Analyzer Operator Manual Assy & TEST, Spare PS38This page intentionally left blank