6.0CALCULATING MOLY CONVERTER EFFICIENCIES

6.1PURPOSE

To provide instructions on how to calculate the efficiency of a Moly converter when using a GPT method of testing converters, by using the US EPA method, where the actual concentration of ozone is not a factor in the accuracy of the calculation of the converter efficiency. This procedure is based on the Code of Federal Regulations, Title 40, Chapter I, subchapter C, Part 50, Appendix F.

6.2TOOLS

API T700 calibrator with O3 Gen option (or equivalent)

Moly Test Data Sheet (provided on page following this procedure)

6.3PARTS

NONE

6.4PROCEDURE

For the purpose of providing an example, this procedure uses 450 PPB NO gas as the reference point, you don’t have to pick these values, they are just an example. There is also an assumption that the analyzer has a good calibration done @ 450 PPB NO span gas. If this is not the case, then once you have finished the leak check on the analyzer, input your 450 PPB NO span gas and calibrate the analyzer.

Note

For the GPT to be performed correctly, there must be a minimum of

 

10% more NO than O3 produced. For example, if the Ozone produced

 

is 400 PPB then the NO used must be 440 PPB or more. Typically 450

 

PPB NO is titrated against 400 PPB of Ozone.

1.Leak check machine to ensure that there are no leaks in the analyzer.

2.For the duration of this test, set the CE factor to 1.0000 (100%) in the instrument firmware (this would be in the CAL-CONC-CONV-SET menu).

07271B DCN6646

35

Image 37
Contents For use with the T200 Operators Manual, P/N Manual AddendumSAN DIEGO, CA Page About this Manual This page intentionally left blank Safety Messages Mise EN Garde Consignes DE SécuritéTable of Contents 07271B DCN6646 Do not pressurize the Sample/Span/Zero port above ambient T201 Ammonia AnalyzerTeledyne API Model T201 NH3 Analyzer Operator Manual T201 and M501 NH 3 Pneumatic Flow Outlined in the following equations Theory of OperationMinimizing PMT Drift Sample Filtration Special Considerations for Ammonia MeasurementT201 Operating Specifications T201 Analyzer SpecificationsThis page intentionally left blank Calibration Procedure Step No Action Comment Zeroing the AnalyzerSpanning the Analyzer with Nitric Oxide GAS No Conv Spanning the Analyzer with Ammonia GAS Efficiency value through the user interface, if required TNx Conv Exit NH3 Calibration Procedure NH3 Gas through the Sample Port NH3 Calibration Procedure NH3 Gas through the Span Port This page intentionally left blank Strip Chart Configurable Analog Output OverviewHyperlink + +Analog Output Pin-Outs M501 NH3 Maintenance Maintenance ScheduleDescription Replacing the Catalytic CartridgeThermocouple Location Replacing the ThermocoupleTeledyne API Model T201 NH3 Analyzer Operator Manual Flow Check Autozero Flow Check This page intentionally left blank Alarm Alarms and CautionsAlarms and Cautions Calculating Moly Converter Efficiencies ProcedurePurpose ToolsCalculating Moly Converter Efficiencies Moly Test Data Sheet This page intentionally left blank To our Website or contact Sales for recent updates T201 Spare Parts ListPCA,TEC DRIVER,PMT,KB Assy & TEST, Spare PS38 Teledyne API Model T201 NH3 Analyzer Operator ManualThis page intentionally left blank